top of page

Measuring Formation Damage Before It Reaches the Well

Writer: Tech Inc
Tech Inc
14 hours ago
2 min read

Every well is drilled with a fluid that the reservoir has never met. Some of that fluid, and the fine solids it carries, enters the rock around the wellbore. If it blocks the pores, the well produces less for the rest of its life. The Tech Inc. Return Permeability Test Apparatus (TI-RPTA-350) measures that harm in the laboratory, before the fluid is chosen.

What the test does

A cylindrical core of reservoir rock is held in a sleeve under confining pressure and heated to reservoir temperature. First, liquid is pumped through it and its permeability is measured. Next, drilling mud is circulated across one face of the core under pressure, as it would flow past the rock in the well; the filtrate that passes through the core is collected and weighed continuously. Finally, liquid is flowed again in the production direction and permeability is measured once more. The second value divided by the first is the return permeability.

Why dynamic filtration matters

In a well, mud moves along the rock face and erodes the filter cake while it forms. A static test cannot reproduce this. Here a magnetically coupled pump recirculates mud from a heated 1,000 cc reservoir through a flow channel across the core face, at up to 6,000 psi (414 bar) and 350 °F (177 °C).

Testing the cure as well as the damage

Acid treatments are used to remove filter cake and restore flow. The apparatus has a separate acid path, lined and with tantalum wetted surfaces, fed from its own lined accumulator. Inhibited hydrochloric acid up to 20% and organic acids can be injected, and the permeability recovered is measured on the same core.

What the laboratory receives

  • Core holder for 1.5 in and 1 in cores up to 6 in long, 7,500 psi (517 bar) confining pressure.

  • Air bath, precision injection pump, back-pressure regulator and confining pressure system.

  • Software with stored test sequences, live trends, alarms and export of the full record.

See the product page and datasheet, or the full range of core analysis instruments. To discuss your test programme, write to mail@techincresearch.com or call +1 (289) 622-1032.

Disclaimer: specifications are typical and can change with the configuration, the fluids used, adherence to the operation manual and other conditions. Images are for representation only. Final data are confirmed in the quotation.