TFF Concentration and Diafiltration Calculator: CF, Diavolumes, Yield and Area

Ultrafiltration and diafiltration steps are planned with four numbers: concentration factor, diavolumes, flux and area. This calculator links them and estimates impurity removal and product yield.
How it works
Concentration factor CF = V₀ ÷ V₁.
Constant-volume diafiltration: impurity remaining = exp(−N × S), with S the sieving coefficient (1 for a freely passing salt).
Product yield: CF^(R − 1) through concentration, exp(−N(1 − R)) through diafiltration, with R the product rejection.
Process time = permeate volume ÷ (flux × area).
Worked example
10 L to 1 L, then 5 diavolumes, salt S = 1, product R = 99%: 99.3% of the salt removed, 93% product yield, 14 L permeate; at 30 LMH on 0.1 m² the step takes 4.7 h, or 0.23 m² finishes in 2 h.
Limits
Flux falls as the product concentrates; use an average measured on a pilot run. Rejection and sieving change with concentration and cross-flow. Our lab UF skids and hollow-fibre module kits give the pilot data. All Tech Inc. calculators and converters.
Related tools: Normalised water permeability, MWCO to pore size, Module area, Vmax filter sizing. All 57 calculators and converters.
Write to mail@techincresearch.com or call +1 (289) 622-1032.
Disclaimer: this calculator and article give estimates from simplified, published equations and the values you enter, for general information only and without warranty. They are not engineering, design, safety or regulatory advice and are no substitute for measurement, the equipment or chemical maker's instructions, the applicable standards or the judgement of a qualified engineer. Do not rely on them for safety-critical decisions. Tech Inc. accepts no liability for loss, damage or injury arising from their use. Equipment specifications are typical and may change; results depend on the membranes and conditions, how the equipment is used and operation in line with the operation manual. See our product information disclaimer and terms of use.