Vmax Filter Sizing Calculator: Filterability Test to Production Filter Area

Writer: Tech Inc
Tech Inc
1 day ago
2 min read

The Vmax method sizes sterilising and clarifying filters from a ten-minute test on a small disc. This calculator does the fit and the scale-up.

How it works

Gradual pore plugging gives t ÷ V = t ÷ Vmax + 1 ÷ Qi, a straight line of t/V against t. Its slope is 1/Vmax and its intercept 1/Qi. Throughput in time t_b per unit area = Vmax (1 − exp(−t_b Qi ÷ Vmax)); the production area is the batch volume divided by that, times a safety factor.

Worked example

3.8 cm² disc at 0.7 bar: 150 mL at 2 min, 300 mL at 5 min, 480 mL at 10 min. Vmax = 1.08 L, Qi = 86 mL/min, so 2,840 L/m². A 500 L batch in 60 min with a 1.5 safety factor needs 0.27 m².

Limits

The model holds when the line is straight; if t/V curves, the fluid is cake-forming and a different model applies. Keep the same pressure, temperature and prefilter at scale. Our dead-end cells, bubble point tester and cartridge test rigs run the tests behind the sizing. All Tech Inc. calculators and converters.

Write to mail@techincresearch.com or call +1 (289) 622-1032.

Disclaimer: this calculator and article give estimates from simplified, published equations and the values you enter, for general information only and without warranty. They are not engineering, design, safety or regulatory advice and are no substitute for measurement, the equipment or chemical maker's instructions, the applicable standards or the judgement of a qualified engineer. Do not rely on them for safety-critical decisions. Tech Inc. accepts no liability for loss, damage or injury arising from their use. Equipment specifications are typical and may change; results depend on the membranes and conditions, how the equipment is used and operation in line with the operation manual. See our product information disclaimer and terms of use.