top of page

Bubble Point and Pore Size Calculator: Convert Pressure to Pore Diameter

Writer: Tech Inc
Tech Inc
1 day ago
2 min read

Choose whether to find pore size from a bubble point, or the bubble point to expect for a pore size. Pick the wetting liquid or enter its surface tension, and the contact angle if the liquid does not wet fully.

How it works

P = 4 γ cos θ ÷ D, rearranged as D = 4 γ cos θ ÷ P

  • γ: surface tension of the wetting liquid

  • θ: contact angle (0° for full wetting)

  • D: diameter of the largest pore, from the first bubble point

Worked example

Water (72.8 mN/m at 20 °C) with a bubble point of 3.0 bar gives a largest pore of about 0.97 µm. A 0.2 µm pore needs about 14.6 bar with water but only about 4.5 bar with ethanol, which is why porometers use low-surface-tension liquids.

Limits

The equation assumes cylindrical pores. ASTM F316 covers membrane filters with maximum pore sizes from 0.1 to 15 µm and notes that the result alone should not be used to state particle retention.

Write to mail@techincresearch.com or call +1 (289) 622-1032.

Disclaimer: this calculator and article give estimates from simplified, published equations and the values you enter, for general information only and without warranty. They are not engineering, design, safety or regulatory advice and are no substitute for measurement, the equipment or chemical maker's instructions, the applicable standards or the judgement of a qualified engineer. Do not rely on them for safety-critical decisions. Tech Inc. accepts no liability for loss, damage or injury arising from their use. Equipment specifications are typical and may change; results depend on the membranes and conditions, how the equipment is used and operation in line with the operation manual.