From Bubble Point to Pore Size: The Capillary Equation, Worked Examples and Why the Wetting Liquid Matters

The bubble point test is one of the oldest ways to size the pores of a membrane or filter: wet the sample, raise gas pressure on one side, and note the pressure at which gas first pushes liquid out of the largest pore. ASTM F316 standardises the bubble point and mean flow pore tests for membrane filters with maximum pore sizes from 0.1 to 15 µm.
The equation
The pressure needed to empty a cylindrical pore is:
P = 4 γ cos θ ÷ D, so D = 4 γ cos θ ÷ P
D: pore diameter.
γ: surface tension of the wetting liquid.
θ: contact angle between liquid and pore wall (0° for a liquid that wets fully, so cos θ = 1).
P: bubble point pressure.
Worked example 1: water
Water, γ ≈ 0.072 N/m, fully wetting, bubble point 3.0 bar (300,000 Pa):
D = 4 × 0.072 ÷ 300,000 = 9.6 × 10⁻⁷ m = 0.96 µm
Worked example 2: why porometers use low-surface-tension liquids
For a 0.2 µm pore:
Water (72 mN/m): P = 4 × 0.072 ÷ 0.2 × 10⁻⁶ = 14.4 bar
Isopropanol (about 22 mN/m): P = 4.4 bar
A lower surface tension brings the test pressure down and helps the liquid wet hydrophobic materials fully, so the cos θ = 1 assumption holds.
What the number means, and what it does not
The first bubble point gives the largest pore. Raising pressure further and measuring gas flow through wet and dry samples gives the mean flow pore and the pore size distribution.
The equation assumes circular pores. ASTM F316 notes the result is an effective limiting diameter and should not be the sole basis for stating particle retention.
Temperature changes surface tension, so record it.
A bubble point far below expected also flags a damaged membrane or a leaking seal.
Equipment
Our capillary flow porometers run bubble point, mean flow pore and full distributions automatically. For filter elements and cartridges, see the Filter Element Bubble Point Tester and the Filter and Membrane Integrity Tester. Background: membrane porometry and membrane porosity.
Write to mail@techincresearch.com or call +1 (289) 622-1032.
Worked examples use illustrative numbers. Equipment specifications are typical and may change; results depend on the membranes and conditions tested, how the equipment is used and operation in line with the operation manual.
Sources
ASTM International, ASTM F316-03(2019) Standard Test Methods for Pore Size Characteristics of Membrane Filters by Bubble Point and Mean Flow Pore Test.
Prof Steven Abbott, Practical Coatings, Pore Size Measurement.