MWCO to Pore Size Calculator: Stokes Radius and Equivalent Pore Diameter

A 10 kDa MWCO is a retention figure, not a pore size. This calculator turns a molecular weight into a solute radius, then into the pore diameter a membrane would need to hold that solute back by size alone.
How it works
Three published routes, because the answer depends on the marker family:
PEG and PEO: water diffusivity from log D = −4.11 − 0.48 log(MW) (D in cm²/s), then the Stokes–Einstein radius r = kT ÷ (6πμD).
Generic solute (Worch): r = 2.0374 × 10⁻¹¹ × MW^0.53 m, the equivalent sphere radius often used to convert pore radius to MWCO.
Globular protein: the minimum radius R = 0.066 × MW^(1/3) nm for a compact sphere; real proteins are 10 to 30% larger.
The equivalent pore diameter is twice the radius.
Worked example
10 kDa gives 2.6 nm (PEG), 2.7 nm (generic) and 1.4 nm (compact protein): pore diameters of about 5.3, 5.4 and 2.8 nm. The spread is why PEG, dextran and protein MWCO values for the same membrane never agree. See MWCO marker molecules by kDa and common MWCO test mistakes.
Limits
Correlations, not measurements. For a measured pore size distribution, use capillary flow porometry on MF and open UF membranes, and a MWCO test with the right marker on tighter membranes. We also run MWCO tests as a service. All Tech Inc. calculators and converters.
Related tools: Bubble point and pore size, Porosity and tortuosity, Hagen–Poiseuille pore flux, TFF and diafiltration. All 57 calculators and converters.
Write to mail@techincresearch.com or call +1 (289) 622-1032.
Disclaimer: this calculator and article give estimates from simplified, published equations and the values you enter, for general information only and without warranty. They are not engineering, design, safety or regulatory advice and are no substitute for measurement, the equipment or chemical maker's instructions, the applicable standards or the judgement of a qualified engineer. Do not rely on them for safety-critical decisions. Tech Inc. accepts no liability for loss, damage or injury arising from their use. Equipment specifications are typical and may change; results depend on the membranes and conditions, how the equipment is used and operation in line with the operation manual. See our product information disclaimer and terms of use.
Sources
Separation and Purification Technology 354 (2025) 128612, Imbrogno et al., Molecular weight cut off (MWCO) determination in ultra- and nanofiltration: review of methods, via KIT (PEG diffusivity correlation; Worch 1993 radius correlation).
Biological Procedures Online 11 (2009) 32, Erickson, Size and Shape of Protein Molecules at the Nanometer Level.