MWCO Marker Molecules by kDa: Which Test Solutes to Use, from 0.1 to 1,000 kDa

Writer: Tech Inc
Tech Inc
2 days ago
3 min read

A molecular weight cut-off (MWCO) test filters solutes of known molecular weight and finds the weight at which the membrane retains 90%. The answer depends heavily on which molecules you use. This table summarises the usual markers for each range. For the full test method, see our MWCO measurement guide.

Bottom line

  • Choose markers that span the expected cut-off, ideally from a few times below to a few times above it.

  • Use neutral, non-adsorbing markers at low concentration.

  • Report which marker family you used. PEG, dextran and protein results for the same membrane will not match.

Marker molecules by range

MWCO range

Typical markers (approximate molecular weight)

Notes

0.1–1 kDa (tight NF)

Glucose (180 Da), sucrose (342 Da), raffinose (504 Da), α-, β- and γ-cyclodextrin (about 973, 1,135 and 1,297 Da), vitamin B12 (1,355 Da), PEG 200–1,000

Compact sugars give more reliable NF results than PEG

0.7–6 kDa (loose NF, tight UF)

PEG oligomers about 0.7–6 kDa; xylo-, fructo- and malto-oligosaccharides

A one-filtration PEG oligomer method covers this range with HPLC and an evaporative light scattering detector

1–10 kDa (tight UF)

PEG 1–10 kDa; maltodextrins 1–10 kDa

Maltodextrins and dextrans are common in makers' own UF tests

10–100 kDa (UF)

PEG 10–35 kDa; dextrans 10–100 kDa; proteins: lysozyme (14.3 kDa), BSA (66.4 kDa)

Proteins are charged; see below

100–1,000 kDa (open UF)

PEO 100 kDa and above; dextrans up to about 1,000 kDa

Large PEO can deposit on open membranes

Above 1,000 kDa

PEO up to about 4,000 kDa

Beyond this, particle or bead challenge tests take over

PEG, dextran or protein?

  • PEG and PEO are cheap, neutral and available from 0.2 kDa (PEG) to 4,000 kDa (PEO). But they are long, flexible chains that slip through pores more easily, and a 2024 review found that PEG overestimates the MWCO of loose NF membranes compared with compact oligosaccharides.

  • Dextrans are branched and widely used for UF. The withdrawn ASTM E1343 method used non-fouling dextrans for flat-sheet UF membranes with cut-offs from 4,500 to 1,000,000 Da.

  • Proteins are compact and relevant to bioprocessing, but they carry charge. A 2024 study found bovine serum albumin (about 69 kDa) retained above 90% by a 300 kDa membrane at low ionic strength, because of electrostatic repulsion rather than size. Use proteins to test performance with proteins, not to state MWCO.

How the markers are analysed

  • Gel permeation (size exclusion) chromatography with a refractive index detector reads a whole PEG or dextran mixture in one run, giving rejection across the full molecular weight range.

  • HPLC with evaporative light scattering separates PEG oligomers one by one.

  • Total organic carbon suits single markers.

  • UV absorbance suits proteins and vitamin B12.

Equipment

The MWCO Testing System runs the filtration under controlled pressure and crossflow. For small samples, use high-pressure stirred cells; for flat-sheet coupons, a crossflow test cell. We also run MWCO tests for clients: see our MWCO testing service.

Before you start, read the common mistakes in MWCO testing.

Write to mail@techincresearch.com or call +1 (289) 622-1032.

Molecular weights are typical values; check the certificate of each marker lot. Equipment specifications may change; results depend on the membranes, markers and conditions, how the equipment is used and operation in line with the operation manual.