Normalised Water Permeability (NWP) Calculator for TFF Cassettes and Modules

Writer: Tech Inc
Tech Inc
1 day ago
2 min read

NWP is the water permeability of a TFF cassette or hollow-fibre module corrected to a reference temperature. Compare it before use, after use and after cleaning to decide whether the membrane is clean enough to reuse.

How it works

NWP = [V ÷ (t × A × TMP)] × TCF, where the temperature correction factor TCF is the ratio of water viscosity at the test temperature to that at 20 °C (or 25 °C). NWP recovery is the post-cleaning NWP divided by the pre-use NWP.

Worked example

500 mL in 2 min through 0.1 m² at 1 bar and 22 °C: 150 LMH/bar at test temperature, TCF 0.953, NWP 143 LMH/bar at 20 °C.

Limits

Use clean water or buffer, a stable TMP and a module fully wetted. Makers set their own acceptance limits for NWP recovery; treat 70 to 80% as a typical threshold rather than a rule. Our clean water flux QC tester automates this test for module QC. All Tech Inc. calculators and converters.

Write to mail@techincresearch.com or call +1 (289) 622-1032.

Disclaimer: this calculator and article give estimates from simplified, published equations and the values you enter, for general information only and without warranty. They are not engineering, design, safety or regulatory advice and are no substitute for measurement, the equipment or chemical maker's instructions, the applicable standards or the judgement of a qualified engineer. Do not rely on them for safety-critical decisions. Tech Inc. accepts no liability for loss, damage or injury arising from their use. Equipment specifications are typical and may change; results depend on the membranes and conditions, how the equipment is used and operation in line with the operation manual. See our product information disclaimer and terms of use.

Sources

Comments


bottom of page