Sample Cylinder Fill Calculator: 80% Fill, Outage Tube Length and the Thermal Expansion Check

A liquid or LPG sample cylinder filled too far becomes liquid-full when it warms, and a liquid-full cylinder has no compressibility to absorb the expansion. This calculator turns the fill rule into numbers for your cylinder and your fluid.
How it works
Liquid volume at sampling = capacity × fill fraction; sample mass = that volume × density at sampling temperature.
Outage = 1 − fill. An outage tube that reaches 20% of the cylinder length down from the top stops the fill at 80%, which is why the tube length is reported.
Expansion: the same mass at the hottest temperature occupies mass ÷ hot density. The calculator reports how full the cylinder is at that temperature and the fill that leaves 5% vapour space.
Worked example
A 500 mL cylinder filled to 80% with a liquid of 510 kg/m³ holds 400 mL and 204 g. At 50 °C the density falls to 470 kg/m³, so the liquid occupies 434 mL, 86.8% of the cylinder; the fill could go to 87.5% before the vapour space shrank to 5%. The outage tube for an 80% fill in a 300 mm cylinder is 60 mm long.
Limits
The expansion check needs the density at the hot temperature; for LPG and NGL use the density from the composition, not an assumed value. Floating-piston (constant-pressure) cylinders keep the sample single-phase without an outage and are the right choice for spot samples that must not flash. Our sample cylinder assemblies, constant-pressure cylinders, RVP sampling cylinders and sample cylinder cleaning station cover the full cycle; the sampling systems range has the panels they connect to. All Tech Inc. calculators and converters.
Related tools: RVP to TVP, Sample line lag time, Gas cylinder contents, Liquid flow units. All 57 calculators and converters.
Write to mail@techincresearch.com or call +1 (289) 622-1032.
Disclaimer: this calculator and article give estimates from simplified, published equations and the values you enter, for general information only and without warranty. They are not engineering, design, safety or regulatory advice and are no substitute for measurement, the equipment or chemical maker's instructions, the applicable standards or the judgement of a qualified engineer. Do not rely on them for safety-critical decisions. Tech Inc. accepts no liability for loss, damage or injury arising from their use. Equipment specifications are typical and may change; results depend on the fluids and conditions, how the equipment is used and operation in line with the operation manual. See our product information disclaimer and terms of use.
Sources
US Code of Federal Regulations, 49 CFR 173.315 Compressed gases in cargo tanks and portable tanks (filling density limits).
ASTM International, ASTM D1265 Standard Practice for Sampling Liquefied Petroleum (LP) Gases, Manual Method.
ASTM International, ASTM D3700 Standard Practice for Obtaining LPG Samples Using a Floating Piston Cylinder.