Orifice Flow Meter Calculator (ISO 5167): Flow from Differential Pressure for Gas and Liquid

The orifice plate is the workhorse of gas and process flow measurement, and its calculation is laid down to the last coefficient in ISO 5167. This calculator implements the Part 2 equations so a metering run, a test rig or a transmitter range can be checked without a flow computer.
How it works
Mass flow q_m = C ε (π/4) d² √(2 ΔP ρ₁) ÷ √(1 − β⁴).
Discharge coefficient C from the Reader-Harris/Gallagher equation as a function of β, pipe Reynolds number and the tapping geometry (corner, flange or D and D/2), solved iteratively because the Reynolds number depends on the flow being sought.
Expansibility ε for gases from the ISO 5167-2 expression with the isentropic exponent; gas density at upstream conditions with a compressibility Z (van der Waals for natural gas unless you enter one).
Checks against the standard's limits of use: pipe diameter 50 to 1,000 mm, β 0.10 to 0.75, minimum Reynolds number.
Worked example
A 102.3 mm line (4-inch schedule 40) with a 55 mm bore and flange taps, 250 mbar of differential pressure, natural gas (molecular weight 17.5) at 40 bar(g) and 30 °C: density 31.2 kg/m³, C = 0.6037, ε = 0.998, Reynolds number 2.1 million, 6,720 kg/h, which is 8,610 Nm³/h or 5,360 SCFM.
Limits
Installation effects (upstream straight length, flow conditioners, plate edge condition and eccentricity) are assumed to meet the standard; they are the dominant source of error in practice. Pulsating flow, wet gas and ΔP above about 25% of the line pressure are outside the method. For custody transfer the flow computer applies the standard with live density and composition, and North American gas uses AGA 3 / API MPMS 14.3 with its own coefficient equation. Our gas metering skids and pressure reducing and metering stations are built with orifice runs to this series; the ISO 5167 guide explains the standard and the gas flow unit converter the reference conditions. All Tech Inc. calculators and converters.
Related tools: Gas flow units, Restriction orifice sizing, Pressure units (gauge/absolute), Control valve Cv and Kv. All 67 calculators and converters.
Write to mail@techincresearch.com or call +1 (289) 622-1032.
Disclaimer: this calculator and article give estimates from simplified, published equations and the values you enter, for general information only and without warranty. They are not engineering, design, safety or regulatory advice and are no substitute for measurement, the equipment or chemical maker's instructions, the applicable standards or the judgement of a qualified engineer. Do not rely on them for safety-critical decisions. Tech Inc. accepts no liability for loss, damage or injury arising from their use. Equipment specifications are typical and may change; results depend on the fluids and conditions, how the equipment is used and operation in line with the operation manual. See our product information disclaimer and terms of use.
Sources
ISO, ISO 5167-2:2022 Measurement of fluid flow by means of pressure differential devices, Part 2: Orifice plates, standard summary page.
Reader-Harris, M., Orifice Plates and Venturi Tubes (Springer, 2015), via Springer.