top of page

TDS Meters vs Conductivity Meters: How They Work and Which to Use

Writer: Tech Inc
Tech Inc
11 hours ago
2 min read

Quick answer: a TDS meter is a conductivity meter that multiplies the conductivity by a conversion factor, usually between 0.5 and 0.7, to estimate total dissolved solids in mg/L. Conductivity (µS/cm) is the measured value and is the better choice for process control, for example on RO permeate. TDS in mg/L is the better choice for reporting water quality, and the gravimetric method is the reference.

What each meter actually measures

Both meters pass an alternating current between electrodes and measure how easily the water carries it. Dissolved ions carry the current, so conductivity rises with salt content. The TDS meter then applies a factor that depends on which salts are present.

Factor basis

Typical factor

Suits

NaCl

about 0.47 to 0.50

RO permeate, seawater-derived water

KCl

about 0.55

Calibration standards

'442' natural water

about 0.65 to 0.70

Surface and ground water

The same water can read 500 or 700 mg/L depending on the factor chosen, so always state the factor with a TDS result. The conductivity to TDS converter shows the effect of each factor.

Temperature compensation

Conductivity rises roughly 2% for every degree Celsius, so readings are referred to 25 °C. Check that the meter compensates automatically and that the coefficient suits the water; very pure water behaves differently from brine.

Choose the right cell constant

Cell constant (K)

Range it suits

Typical use

0.1 cm⁻¹

Below about 200 µS/cm

RO permeate, boiler feed, pure water

1.0 cm⁻¹

About 10 to 20,000 µS/cm

Drinking water, process water

10 cm⁻¹

Above about 1,000 µS/cm

Brine, RO concentrate, seawater

Accuracy and calibration

  • Calibrate with a standard close to your samples, not just one high standard.

  • Rinse the cell with sample before reading, and keep the electrodes clean.

  • For pure water, measure in a flow cell to avoid carbon dioxide pick-up from air.

  • For legal or contractual TDS figures, use the gravimetric method (filter, evaporate and dry at 180 °C).

Which one should you use?

Task

Better choice

Why

Watching RO permeate quality

Conductivity

Direct, fast and independent of any factor

Reporting drinking-water quality

TDS (mg/L)

Regulations are written in mg/L

Membrane rejection tests

Conductivity

Rejection is a ratio, so the factor cancels

Field screening of wells

Either

Note the factor if TDS is used

Tech Inc.'s handheld conductivity meter reads conductivity, TDS and temperature, and its membrane test systems log feed and permeate conductivity continuously. For methods and interpretation, see how to test TDS in water.

Frequently asked questions

Is a TDS meter accurate?

It is as accurate as its conductivity reading and the factor chosen. With the wrong factor it can be off by 20% or more.

What is a good TDS for drinking water?

Many guidelines treat below about 500 to 600 mg/L as good for taste; local regulations set the limit.

Why does my TDS meter read zero on RO water?

Very pure water is outside the range of a meter with a 1.0 cell constant. Use a 0.1 cell.

Do TDS meters detect bacteria or organics?

No. They respond only to dissolved ions.

Values in this guide are typical and for general information. Limits differ by membrane, instrument and chemical; always follow the manufacturer's data and your site procedures. Equipment specifications are confirmed in the quotation.

Comments


bottom of page