Dew Point Meters: Measuring Moisture in Gas and Compressed Air

Quick answer: a dew point meter measures the temperature at which water vapour in a gas would start to condense. The lower the dew point, the drier the gas. Always state whether a value is a pressure dew point (at line pressure) or an atmospheric dew point, because the same gas has a much lower dew point at atmospheric pressure.
Why dew point matters
Compressed air: water causes corrosion, freezing and contamination of products and instruments.
Natural gas: water forms hydrates and corrodes pipelines, so contracts limit water content.
Process and inert gases: moisture spoils catalysts, analysers and sensitive products.
Sensor types
Sensor | Principle | Strengths | Watch-outs |
Capacitive polymer or aluminium oxide | Capacitance changes as the film absorbs water | Compact, inline, low cost | Needs periodic calibration; drift at very low dew point |
Chilled mirror | Cools a mirror until dew forms | Fundamental reference method | Slower, needs a clean mirror |
Quartz crystal microbalance | Frequency shift as water is absorbed | Fast and sensitive at low levels | Higher cost |
Tunable diode laser (TDLAS) | Infrared absorption by water | Natural gas, no contact with the gas | Higher cost, specific to the gas |
Compressed air classes
ISO 8573-1 sets humidity classes for compressed air by pressure dew point:
Class | Pressure dew point |
1 | −70 °C or lower |
2 | −40 °C or lower |
3 | −20 °C or lower |
4 | +3 °C or lower |
5 | +7 °C or lower |
6 | +10 °C or lower |
Installing a dew point sensor
Measure at line pressure if you need pressure dew point, or let down through a regulator for atmospheric readings.
Use a sample cell and short stainless steel tubing; avoid plastic tubing, which releases moisture.
Keep a small continuous purge flow past the sensor.
Protect the sensor from liquid water and oil with a filter upstream.
Allow time to dry down: a wet sensor can take hours to reach a very low dew point.
Calibrate at the interval the maker recommends, often yearly.
Convert dew point to ppmv, lb/MMscf or mg/Nm³ with the dew point to ppm converter. Gas conditioning skids compared covers drying and filtration ahead of analysers, and Tech Inc.'s analyser systems bring sampling, conditioning and moisture measurement together on one skid, for example the sampling cabinet for H₂S and moisture analysers.
Frequently asked questions
What is the difference between dew point and relative humidity?
Relative humidity depends on temperature; dew point is an absolute measure of how much water is in the gas.
Why is atmospheric dew point lower than pressure dew point?
Expanding the gas lowers the partial pressure of water, so it would condense only at a lower temperature.
How dry should instrument air be?
A pressure dew point at least 10 °C below the lowest temperature the air will see is a common rule; many sites specify −40 °C.
Can a dew point sensor be left in the line?
Yes, inline sensors are made for continuous monitoring, protected by a filter.
Values in this guide are typical and for general information. Limits differ by membrane, instrument and chemical; always follow the manufacturer's data and your site procedures. Equipment specifications are confirmed in the quotation.