ISO 10715 Explained: Natural Gas Sampling Guidelines and What They Mean for a Sampling System

A gas chromatograph is only as good as the sample it receives. ISO 10715 is the international guideline for taking that sample from a natural gas stream, and most custody-transfer and quality contracts cite it alongside GPA 2166 and API MPMS Chapter 14.1. The current edition is ISO 10715:2022, which replaced the 1997 edition.
What it covers
According to ISO's published scope, the standard gives guidance on sampling natural gas and similar gaseous fuels for subsequent analysis: spot (grab) sampling, composite (incremental) sampling and continuous (direct) sampling to an on-line analyser. It discusses the sampling point and probe, the transfer line, the sample container, and the sources of error that distort a sample: condensation of heavy ends, adsorption and desorption on surfaces, contamination from a previous sample, and leakage. Hydrocarbon dew point relative to the sample temperature runs through the whole document, because a sample that cools below its dew point at any point loses the components that set its calorific value.
What it means in practice
Probe placement: a probe into the centre third of the pipe, away from fittings and liquid traps; the standard explains why a wall tap gives an unrepresentative sample.
Keep it above the dew point: heated probes, heat-traced lines and heated regulators are the consequence; the Joule–Thomson cooling across a regulator is the usual place a sample goes two-phase.
Materials and volume: small internal volume, materials that do not adsorb or react (passivated or sulfinert-type surfaces for H₂S and mercaptan analysis), and purge times long enough to clear the line.
Cylinders: constant-pressure (piston) cylinders keep the sample single-phase; fill methods are chosen to avoid fractionation.
How this shapes our equipment
Our gas sampling panels for process, natural and sour gas, GC sampling cabinets, constant-pressure sample cylinders and heated sample probes are configured around these points: probe depth, heat tracing from the tap to the cylinder or analyser, low-volume sulfinert tubing for sulphur species, and purge-to-vent sequencing. The sample line lag time calculator and dew point converter do the numbers that the guideline asks for. All our standards guides.
Send your application, materials and the standards your contract names to mail@techincresearch.com or call +1 (289) 622-1032, and we will return a configuration, datasheet and quotation.
This article is our commentary on the standard's published scope, not a substitute for the standard itself; always work to the current edition named in your contract. Equipment specifications may change; performance depends on the application, how the equipment is used and operation in line with the operation manual.