Application Note: Hydrogen, Nitrogen and Zero Air from Generators in the Laboratory

At a glance. A gas chromatograph needs hydrogen, nitrogen and zero air; a hydrogenation lab needs hydrogen at reactor pressure; a sample-preparation lab needs nitrogen to blow down extracts before LC-MS or GC-MS. Generators make each gas on demand from water and air, so the lab stores almost no gas and never waits for a delivery.
Gas chromatographs
A three-in-one generator makes hydrogen in a PEM cell, nitrogen with a membrane or PSA module, and zero air with a catalytic purifier, each with its own regulator and alarm. Size it from the flows of every detector and carrier on every GC it feeds, plus a margin.
Hydrogenation
A PEM generator delivers dry hydrogen at moderate pressure; a booster and small buffer vessel reach reactor pressures. Because only a small volume is stored, the hydrogen inventory in the room stays low. The installation still follows the hydrogen rules that apply locally, such as NFPA 2 in many jurisdictions: ventilation, leak detection, and shut-down on alarm.
Sample concentration
A nitrogen evaporator blows a gentle stream of nitrogen onto the surface of each sample in a heated block, concentrating extracts for trace analysis, including PFAS methods. One nitrogen generator sized for both can supply the LC-MS and the evaporator.
Sizing checklist
List each user and its flow and pressure.
Add the peak, not the average, where users start together.
Check purity needed at each user.
Plan water supply, power, ventilation and alarms.
Where Tech Inc. equipment fits
Tech Inc. builds three-in-one laboratory gas generators, high-pressure laboratory hydrogen generators, nitrogen evaporators, LC-MS nitrogen generators and GC gas stations.
Related reading
Sources
NFPA 2, Hydrogen Technologies Code (2023 edition) – National Fire Protection Association, via ANSI
This note is general information. Generators are sized from the laboratory's instruments, flows and site rules.