Application Note: Helium Recovery for NMR, MRI and Cryostat Laboratories

Writer: Tech Inc
Tech Inc
5 days ago
2 min read

At a glance. NMR spectrometers, some MRI magnets, cryostats and low-temperature physics experiments are cooled with liquid helium, which slowly boils off. One litre of liquid becomes roughly 0.75 m³ of gas at normal conditions. Venting it wastes a scarce resource; recovering it protects research from price swings and supply gaps such as the 2026 shortage.

A recovery system

  1. Collection. Boil-off from each magnet or cryostat is piped through low-pressure return lines with flow meters to a gas bag or buffer vessel.

  2. Compression. When the bag fills, a compressor raises the pressure and an after-cooler removes heat and water.

  3. Purification. An adsorption, membrane or cryogenic purifier removes air, moisture and oil; an analyser confirms the purity.

  4. Storage and reuse. The clean helium is stored in cylinders or a high-pressure bank, then sent to a small liquefier or back to the gas supplier for credit.

Making it pay

  • Capture rate. Well-sealed return lines capture most of the boil-off; leaks and open transfers are where gas is lost.

  • Purity. The purer the returned gas, the simpler the purifier and the easier the reliquefaction.

  • Scale. A single NMR lab may only justify gas collection and cylinder return; a campus with several magnets and cryostats can justify a full recovery and liquefaction plant.

  • Operation. The system should run unattended, with alarms to the lab's monitoring system.

Leak testing and other users

Helium leak-test stations and gas pycnometers also use helium that can be reclaimed, often with a higher air content that the purifier has to deal with.

Where Tech Inc. equipment fits

Related reading

Sources

This note is general information. Each system is sized from a survey of the site's helium use.