Gas Pycnometry: Measuring True Density with Helium

At a glance. A gas pycnometer fills a reference chamber with helium, expands it into the sample chamber and measures both pressures. Boyle's law gives the volume the gas cannot enter, which is the sample's solid volume; with the mass, that is its true (skeletal) density. The Tech Inc. Gas Pycnometer automates the cycle and averages repeat runs.
Why helium
Helium atoms are small and do not adsorb at room temperature, so they reach fine pores and give the true skeletal volume. Nitrogen suits materials that helium would diffuse into slowly, and some standards call for it.
Getting repeatable results
Fill the cup: a sample volume at least two-thirds of the cup reduces relative error.
Dry the sample: water vapour adds pressure and lowers apparent density.
Purge: several purge cycles remove air and moisture.
Temperature: keep it steady; thermostatted chambers help with small volumes.
Calibrate: check with the reference sphere regularly.
Where it is used
Industry | Use |
Pharmaceuticals | true density of APIs and excipients for tableting |
Batteries | electrode powders and porosity of coated electrodes |
Additive manufacturing | metal powder density and part porosity |
Ceramics and cement | raw materials and fired parts |
Oil and gas | grain density of core plugs |
Frequently asked questions
True density or bulk density? Pycnometry gives true (skeletal) density. Bulk density includes the spaces between particles.
Can it give porosity? Yes: porosity = 1 − bulk density ÷ true density.
Send your materials to mail@techincresearch.com for a recommendation.