Pressure Gauge Calibration: Procedure and Automation

At a glance. Calibration compares the gauge under test with a more accurate reference at several points rising and falling, and records the errors. The Tech Inc. Automated Pressure Gauge Calibration System sets each pressure automatically and produces the certificate.
Procedure
Exercise the gauge to full scale a few times.
Choose points, often 0, 25, 50, 75 and 100% of range, rising then falling.
Let each point stabilise; tap analogue gauges lightly if the procedure allows.
Record readings and calculate error and hysteresis.
Compare with the gauge's accuracy class; adjust or reject.
Reference accuracy
The reference should be several times more accurate than the gauge under test; a ratio of 4:1 is a common rule.
Frequently asked questions
How often? Set by your quality system and the gauge's duty; critical gauges are often checked every 6 to 12 months.
Transmitters and switches? Yes: the system reads mA outputs and switch points as well as dials.