Why Two Labs Get Different SDI on the Same Water: Filters, Temperature and Technique

Plant operators often see a contractor's SDI differ from their own on the same water. Usually neither is careless: the method has known sources of variation, and ASTM D4189 itself names two of them.
What the standard says
ASTM D4189-23 notes that SDI can vary with water temperature and with the membrane filter manufacturer, so results under different conditions or with different filters may not be comparable.
What research measured
A study by Alhadidi and co-workers (2012) quantified the errors:
Filter resistance had the biggest effect. Within the variation the standard allows, it could shift an SDI of 3 by 2.29 to 3.98.
Timing mattered much less: a one-second error in tf changed SDI by about ±0.07 at SDI 3.
Total error at SDI 3 was estimated at about ±2.11 in the field and ±0.4 in the laboratory.
In other words, two correctly run tests can disagree by more than the difference between "good" and "marginal" water.
Temperature
Water is more viscous when cold, so the same filter passes 500 mL more slowly. SDI is not corrected for temperature, so a winter and a summer SDI of the same water are not strictly comparable. Record temperature with every test.
Practical fixes
One filter source and one lot. Buy 0.45 µm membrane filters in quantity from one supplier and keep lot numbers with results.
Check the filters. Measure ti with clean water on a sample from each lot; a lot that runs much faster or slower than usual will shift results.
Hold pressure. Use a regulator that holds 207 kPa steadily and a calibrated gauge.
Wet and seat the filter correctly and remove trapped air from the holder.
Log temperature, filter lot and operator with every result.
Automate. An online analyser such as the Silton 4750 Max applies the same pressure, timing and calculation every time, which removes much of the operator variation.
Better indices
The same research group proposed ways to normalise SDI for temperature, pressure and filter resistance; see beyond SDI.
Write to mail@techincresearch.com or call +1 (289) 622-1032.
Equipment specifications are typical and may change. SDI results depend on the water, the filters, how the equipment is used and operation in line with the operation manual.
Sources
ASTM International, ASTM D4189-23 Standard Test Method for Silt Density Index (SDI) of Water.
Desalination and Water Treatment 40 (2012) 100–117, Alhadidi et al., Sensitivity of SDI for experimental errors.





Comments