top of page

Application Note: Keeping Light Ends and Dissolved Gas in the Sample

Writer: Tech Inc
Tech Inc
5 days ago
2 min read

At a glance. A vapour-pressure result falls if butanes and pentanes escape from the sample; a dissolved-gas result for a transformer changes if gas leaves the oil or air gets in. Both are solved the same way: a closed container, filled without a gas cap, and carried to the laboratory without opening it.

Vapour pressure of crude, condensate and gasoline

Vapour pressure decides how crude and gasoline are blended, stored, shipped and classified for transport. Open bottles lose light ends, so samples are taken into closed cylinders. ASTM D8009 describes manual piston-cylinder sampling of volatile crude oils, condensates and liquid products from low-pressure points, to prevent loss of light ends before vapour-pressure and flash-point tests.

  • Floating-piston cylinders hold the sample under pressure behind a piston, with no gas cap.

  • Fixed-volume cylinders with an outage tube suit liquids that are cooled before sampling.

  • Seals are chosen for the product: FKM or FFKM for aromatic and sour liquids.

Dissolved-gas analysis of transformer oil

Faults in transformers make gases that dissolve in the oil; their amounts and ratios point to overheating, partial discharge or arcing. IEC 60567 covers sampling of free gases and oil from electrical equipment and the analysis of dissolved gases. Gas-tight glass syringes with three-way stopcocks take oil without air contact or headspace; bottles are used for moisture and other oil tests; a padded, dark case protects samples in transit.

Good practice for both

  1. Flush the sample point to clear dead volume.

  2. Fill without a gas cap and close valves or stopcocks at once.

  3. Label with the point, time, temperature and pressure.

  4. Keep samples cool and test them promptly.

  5. Clean and check containers between uses.

Where Tech Inc. equipment fits

Related reading

Sources

This note is general information. Follow the test method and the laboratory's instructions for each sample.

Comments


bottom of page