UKAS and ISO/IEC 17025 Explained: What an Accredited Test or Calibration Certificate Means

Two words are used interchangeably and mean different things. A company is certified to a management system standard such as ISO 9001. A laboratory is accredited to ISO/IEC 17025 by a national accreditation body, and only that accreditation says its test or calibration results are technically valid.
ISO/IEC 17025
ISO/IEC 17025:2017, General requirements for the competence of testing and calibration laboratories, specifies what a laboratory must demonstrate: impartiality and confidentiality; structural requirements; competent personnel, suitable facilities and equipment; metrological traceability of measurements to the SI through an unbroken chain of calibrations with stated uncertainties; validated methods; handling of samples and records; evaluation of measurement uncertainty; reporting; and a management system. It is written for laboratories, not for manufacturers, and it is assessed, not certified.
UKAS and the other accreditation bodies
UKAS is the United Kingdom's national accreditation body; its counterparts include DAkkS (Germany), COFRAC (France), A2LA and ANAB (USA), NABL (India), EIAC and GAC (Gulf states) and SCC (Canada). They are linked by the ILAC mutual recognition arrangement, so an accredited certificate from one is accepted by the others. Accreditation is for a defined scope: specific tests or calibrations, ranges and uncertainties, listed on the body's website, which is where a certificate should be checked.
Reading a certificate
An accredited calibration certificate carries the accreditation body's symbol with the laboratory's number, states the traceability, gives the measured values with their uncertainty and the conditions, and is signed by an authorised person. A certificate from an ISO 9001-certified company without an accreditation symbol is a manufacturer's declaration; useful, but not the same evidence. Test certificates for equipment (pressure gauges, transmitters, relief valves, flow meters, analysers) are what a client's inspector will ask for at FAT.
What it means for our equipment
Instruments supplied with our systems can be delivered with accredited calibration certificates from laboratories accredited under the ILAC arrangement, and we issue a test certificate for every system we build; see Request a Test Certificate and our testing and quality page. The pressure gauge calibration systems in our range are the equipment behind many of those certificates. All our standards guides.
Send your application, materials and the standards your contract names to mail@techincresearch.com or call +1 (289) 622-1032, and we will return a configuration, datasheet and quotation.
This article is our commentary on the standard's published scope, not a substitute for the standard itself; always work to the current edition named in your contract. Equipment specifications may change; performance depends on the application, how the equipment is used and operation in line with the operation manual.
Sources
ISO, ISO/IEC 17025:2017 General requirements for the competence of testing and calibration laboratories, standard summary page.
UKAS, United Kingdom Accreditation Service, official site.
ILAC, International Laboratory Accreditation Cooperation, official site.