top of page

RO Recovery, Concentration Factor and Mass Balance: The Three Calculations Every Test Needs

Writer: Tech Inc
Tech Inc
1 day ago
2 min read

Every RO test, from a flat-sheet coupon to a pilot skid, rests on three simple calculations. Get them right and the rest of the data makes sense; skip them and errors slip through.

1. Recovery

Recovery, r = Qp ÷ Qf, where Qp is permeate flow and Qf is feed flow. ASTM D4516 calls it conversion and lists it among the operating conditions that change permeate flow and salt passage.

2. Mass balance

What goes in must come out:

Qf × Cf = Qp × Cp + Qc × Cc

where C is concentration and Qc = Qf − Qp is the concentrate flow.

3. Concentration factor

CF = Cc ÷ Cf. With near-full rejection, CF ≈ 1 ÷ (1 − r).

Worked example (illustrative numbers)

  • Feed 100 L/h at 2,000 mg/L; permeate 75 L/h at 40 mg/L.

  • Recovery = 75 ÷ 100 = 75%.

  • Concentrate flow = 25 L/h.

  • Cc = (100 × 2,000 − 75 × 40) ÷ 25 = 7,880 mg/L.

  • CF = 7,880 ÷ 2,000 = 3.94, close to 1 ÷ (1 − 0.75) = 4.

Now measure the concentrate. If it reads far from 7,880 mg/L, something is wrong: a flow meter, a sample, a leak or scaling inside the module.

Lab-test traps

  • Recirculating tank. In many lab rigs the concentrate returns to the feed tank, so the feed gets saltier during the test. Log feed conductivity over time, or run with permeate returned to the tank to keep it steady.

  • Recovery per pass vs overall. A coupon test runs at very low recovery per pass; a plant runs high overall. Compare like with like.

  • Scaling. Higher recovery raises concentrate salts towards their solubility limits. That, not the membrane, often sets the maximum recovery. A high-recovery CCRO test on the real water finds the limit.

Why osmotic pressure follows

A concentration factor of 4 means roughly four times the feed osmotic pressure at the concentrate end. See osmotic pressure calculations.

Our CCRO pilot test skid and CCRO lab test skid log every flow and conductivity for these calculations.

Write to mail@techincresearch.com or call +1 (289) 622-1032.

Worked examples use illustrative numbers. Equipment specifications are typical and may change; results depend on the membranes and conditions tested, how the equipment is used and operation in line with the operation manual.

Sources

Comments


bottom of page