ASTM F316 Explained: Bubble Point and Mean Flow Pore Size of Membrane Filters

Pore size is the first question about any membrane or filter medium. The reference method for measuring it by gas–liquid displacement is ASTM F316, Standard Test Methods for Pore Size Characteristics of Membrane Filters by Bubble Point and Mean Flow Pore Test. The current edition is F316-03(2019).
What it covers
ASTM's published scope covers two pore size properties of membrane filters with maximum pore sizes from 0.1 to 15.0 µm:
Test Method A: the maximum limiting pore diameter of non-fibrous membranes, through the bubble point.
Test Method B: the relative abundance of a specified pore size.
The scope notes that the bubble point can also reveal a damaged membrane, ineffective seals or a system leak, that the effective pore size assumes circular capillaries, and that results should not be the sole basis for stating particle retention.
Our commentary: how the test works in practice
The sample is wetted with a liquid that fills every pore.
Gas pressure rises on one side; the first steady gas flow marks the bubble point, the largest pore.
Pressure keeps rising while gas flow through the wet sample is recorded (the wet curve), then the same is done on the dry sample (the dry curve).
Where the wet curve reaches half the dry curve is the mean flow pore; the shape of the curves gives the pore size distribution.
Pressure converts to pore diameter through the capillary equation; try our bubble point and pore size calculator.
Common mistakes
Incomplete wetting gives a falsely high bubble point.
Evaporation of the wetting liquid during the run shifts the wet curve.
Edge leaks in the holder look like large pores.
Wrong liquid surface tension in the software scales every diameter wrongly.
Tech Inc. equipment
Our capillary flow porometers, including the AL500 EFM and AL1000, measure bubble point, mean flow pore and through-pore distribution by the gas–liquid displacement method of ASTM F316. For filter elements, see the Filter Element Bubble Point Tester (ISO 2942).
Send your line data, product and the standards your contract names to mail@techincresearch.com or call +1 (289) 622-1032, and we will return a configuration, datasheet and quotation.
This article is our commentary on the standard's published scope, not a substitute for the standard itself; always work to the current edition. Equipment specifications may change; performance depends on the application, how the equipment is used and operation in line with the operation manual.
Sources
ASTM International, ASTM F316-03(2019) Standard Test Methods for Pore Size Characteristics of Membrane Filters by Bubble Point and Mean Flow Pore Test, standard summary page.