UF/DF Skids: How Ultrafiltration–Diafiltration Systems Work and How to Specify One

At a glance. A UF/DF skid uses tangential flow filtration (TFF) to do two jobs: ultrafiltration (UF) concentrates the product by removing liquid through the membrane, and diafiltration (DF) washes small solutes out by adding new buffer as permeate leaves. A low-shear pump circulates the feed across flat-sheet cassettes or hollow-fibre modules while the skid holds transmembrane pressure (TMP) and adds buffer to keep volume constant. In constant-volume diafiltration, about 5 diavolumes remove roughly 99% of a freely passing solute and about 7 diavolumes remove roughly 99.9%. Specify the skid around your product, batch volume, target concentration, membrane type and the level of automation and documentation you need.
What is a UF/DF skid?
A UF/DF skid is a self-contained, frame-mounted system that carries the feed tank, recirculation pump, membrane holder or modules, valves, sensors and controls for tangential flow filtration. It is used wherever a biological product must be concentrated and moved into a new buffer without damaging it: monoclonal antibodies and other proteins, vaccines, viral vectors, plasmid DNA, enzymes, and many food and fine-chemical streams.
In TFF the feed flows along the membrane rather than straight into it. The sweeping flow limits the build-up of a gel or cake layer, so the membrane keeps filtering through long runs. The membrane cut-off (MWCO) is chosen so that the product is retained and water, salts and small molecules pass. As a rule of thumb, an MWCO of one third to one fifth of the product's molecular weight is a common starting point; see how to measure membrane MWCO.
The UF/DF process step by step
Flush and equilibrate. Remove storage solution and equilibrate the membrane with buffer; measure clean water flux as a baseline.
Ultrafiltration 1 (concentration). Recirculate the feed and draw permeate until the target volume or concentration is reached.
Diafiltration. Add the new buffer at the same rate as permeate is removed (constant volume) for the required number of diavolumes.
Ultrafiltration 2 (final concentration). Concentrate to the final target, allowing for hold-up volume.
Product recovery. Drain and buffer-flush the retentate side to recover product from the loop and modules.
Clean, test and store. Clean in place, re-measure water flux to confirm recovery, and integrity-test where required.
How many diavolumes do you need?
For constant-volume diafiltration of a solute that passes the membrane freely, the fraction remaining after N diavolumes is approximately:
Fraction remaining = e^(−N)
Diavolumes (N) | Fraction remaining | Removal |
3 | 5.0% | 95.0% |
5 | 0.7% | 99.3% |
7 | 0.09% | 99.9% |
10 | 0.005% | 99.995% |
Solutes that are partly retained, or that bind to the product, need more diavolumes. Running DF at a higher product concentration uses less buffer and less time, until viscosity and flux decline; the optimum is usually found experimentally on a lab skid.
Sizing membrane area
Membrane area is set by the permeate volume, the average flux and the process time:
Area (m²) = permeate volume (L) ÷ [average flux (LMH) × process time (h)]
Measure flux against TMP and concentration on the real feed at small scale, choose an operating point just below the TMP where flux stops rising, and add a safety margin for fouling and batch variation. Our guide to membrane flux and permeability formulas shows the calculations.
What a good UF/DF skid carries
Area | What to look for |
Pump | Low-shear (quaternary diaphragm, rotary lobe or magnetically levitated centrifugal), with turndown for small and large modules |
Membrane format | Flat-sheet cassettes and/or hollow-fibre modules, with holders sized for the area range |
Pressure control | Feed, retentate and permeate transmitters; automatic retentate valve to hold TMP |
Flow | Feed (crossflow) and permeate flow measurement; permeate control for constant-flux operation |
Diafiltration | Buffer pump with level or weight control for constant-volume DF |
Process sensors | Temperature, conductivity and pH; UV where product concentration must be tracked |
Tankage | Jacketed feed tank with low hold-up and good mixing at the final volume |
Hygiene | 316L wetted parts, sanitary fittings, drainable design, CIP and sanitisation; single-use flow paths as an option |
Software | Recipes for UF1/DF/UF2, alarms, trends and batch reports; audit trail and access control for GMP work |
Lab, pilot and production skids
Lab UF/DF systems work with tens of millilitres to a few litres and small cassettes, for feasibility, MWCO selection and DF optimisation.
Pilot skids run tens to hundreds of litres with the same membrane chemistry and path length as production, so that flux and TMP data scale linearly by area.
Production skids add redundancy, full automation and the documentation package (design qualification, FAT/SAT, IQ/OQ) required by the plant.
Keeping the membrane type, channel geometry and crossflow per unit area constant across scales is what makes scale-up predictable.
Specification checklist
Product, molecular weight, concentration range and shear sensitivity.
Batch volume, final volume and acceptable hold-up volume.
Target concentration factor and number of diavolumes.
Membrane format (cassette or hollow fibre), MWCO and area range.
Process time limit and temperature control.
Instruments and control loops required (TMP, constant flux, constant volume DF).
Materials, surface finish, CIP/SIP or single-use, and drainability.
Data integrity and documentation level (R&D, pilot or GMP).
Utilities and footprint.
Tech Inc. builds the tangential flow filtration and diafiltration system and the lab UF skid for process development and pilot work. Send your batch volume, product and target concentration to mail@techincresearch.com for a sized proposal.
Frequently asked questions
What does UF/DF stand for?
Ultrafiltration/diafiltration: concentration by ultrafiltration, then buffer exchange by diafiltration, usually on the same tangential flow filtration skid.
What is the difference between TFF and UF/DF?
TFF is the flow mode (feed sweeps across the membrane). UF/DF is the process carried out with it.
Hollow fibre or cassettes for UF/DF?
Cassettes give high packing density and are standard for proteins; hollow fibres have open channels and lower shear, which suits viral vectors, cells and viscous feeds. See hollow fiber vs flat sheet membranes.
How is a UF/DF skid cleaned?
By clean-in-place with caustic and, where needed, acid or enzyme cleaners, followed by a water flux check against the baseline. See membrane CIP cleaning.