The Time-Lag Method: Measuring CO2 Permeability, Diffusivity and Solubility of Membrane Films

Writer: Tech Inc
Tech Inc
2 days ago
2 min read

Every new CO2-selective polymer or mixed-matrix film starts in the same place: a small disc in a permeation cell. The Tech Inc. Time-Lag Gas Permeation Apparatus measures that disc with the constant-volume, variable-pressure method, the time-lag method.

How the method works

  1. The film is sealed in the cell and both sides are evacuated until the downstream pressure stops rising.

  2. Gas is admitted to the feed side at a set pressure.

  3. A capacitance gauge records the pressure rise in a fixed, calibrated volume on the downstream side.

At first nothing comes through while the gas dissolves into the film and diffuses across it. Then the pressure rises in a straight line.

  • The slope of the straight line gives the permeability.

  • Extending the line back to the time axis gives the time lag, and from it the diffusivity.

  • Solubility follows from the two, because permeability is diffusivity times solubility.

Running CO2, N2, CH4 and H2 in turn gives the ideal selectivity of the film for each pair.

What decides the quality of the data

  • Leak rate. The downstream volume must hold vacuum, so every run starts with a leak check and the leak rate is subtracted.

  • Volume calibration. Permeability is only as good as the downstream volume figure, which is calibrated at the factory.

  • Temperature. Permeability changes with temperature, so the cell sits in an oven.

The standards

ASTM D1434 sets out the manometric method for gas permeability of plastic film and sheeting, and ISO 15105-1 the differential-pressure methods for gas-transmission rate. The apparatus follows both.

What comes next

Pure-gas data rank films quickly, but they can flatter a membrane. The next step is mixed-gas testing; see our guide to mixed-gas data and the Flue-Gas CO2 Membrane Test System. All our gas separation rigs are on Membrane Testing Systems.

What to send us

The film types and sizes, the gases, the temperature range and your site utilities.