Sampling Condensate, NGL and LPG Without Losing Light Ends: Floating-Piston Cylinders Explained

Writer: Tech Inc
Tech Inc
Nov 7
5 min read

Condensate, NGL and LPG are only liquid because they are under pressure. Take a sample into an open container, or let the pressure fall in a partly filled cylinder, and the lightest components boil off first. The sample that reaches the gas chromatograph is then heavier than the product in the line, and every property calculated from it, from vapour pressure to heating value, is biased. This guide is for laboratory, metering and process engineers who need pressurised hydrocarbon samples they can defend.

Why light ends are lost

A pressurised liquid sits above its bubble point. If the pressure falls below that point, vapour forms, and the vapour is rich in methane, ethane and propane. Any vapour that is vented, or that stays in the cylinder's head space and is not recombined, changes the composition. Common causes are filling a cylinder that already contains air or a different gas, opening a valve too quickly so the liquid flashes across it, and letting a cylinder warm up with no room for the liquid to expand.

The effect is largest for the lightest products. A small change in the measured ethane or propane content of an NGL stream changes its calculated vapour pressure and value, and for LPG it can put a sample out of specification on paper when the product in the tank is fine.

Two cylinder types, two methods

Fixed-volume cylinders with an outage tube. ASTM D1265 describes the manual sampling of LPG into single- or double-valve cylinders. After filling, liquid is vented through an outage tube to leave a vapour space, so the cylinder is never liquid-full when it warms. The method is simple and widely used for LPG quality tests, but the venting step and the vapour space mean the sample is not an exact copy of the line composition, which matters for compositional analysis of lighter mixtures.

Constant-pressure, floating-piston cylinders. ASTM D3700 covers obtaining LPG samples with a floating-piston cylinder, and GPA 2174 covers obtaining liquid hydrocarbon samples for analysis by gas chromatography, including floating-piston methods. A free piston divides the cylinder: sample on one side, an inert precharge gas on the other. The sample enters against the precharge, stays above its vapour pressure throughout and never forms a vapour space. Nothing is vented from the sample side.

Factor

Outage-tube cylinder (ASTM D1265)

Floating-piston cylinder (ASTM D3700, GPA 2174)

Sample phase

Liquid with a vapour space

Single-phase liquid at pressure

Venting

Liquid vented through the outage tube

No venting from the sample side

Light-end retention

Adequate for many LPG quality tests

Suited to compositional analysis

Cost and handling

Lower; simple valves

Higher; piston, seals and precharge

Typical use

LPG specification testing

Condensate, NGL and LPG composition; custody transfer

Setting the precharge

The precharge pressure is the heart of the method. It must be above the vapour pressure of the sample at the highest temperature the cylinder will see, so the sample stays liquid, and below the line pressure, so the sample can push the piston and fill the cylinder. Inert gases such as nitrogen, helium or argon are typical. The laboratory should confirm which gas to use, because it can interfere with some analyses if it ever leaks past the piston. Check the piston seals and the precharge before every use.

Fill limits and mixing

A floating-piston cylinder must not be filled completely. Room is left on the precharge side so the liquid can expand if the cylinder warms during transport, and the maximum fill is set by the cylinder maker and the sampling standard; follow their figure, not a rule of thumb. Most cylinders show piston position with an indicator rod or a magnetic follower. Before analysis the sample is mixed, typically with a mixing ball or a rocking device on the sample side, so any components that separated in transit are recombined.

Tech Inc. LPG, LNG and hydrogen sampling system with company logo

Spot samples versus automatic pressurised sampling

Spot samples, taken by connecting a cylinder to a sample point through a probe, are fine for process checks and quality control. For custody transfer of condensate, NGL or LPG, a single spot sample does not represent a batch whose composition changes over hours. An automatic pressurised sampling system takes flow-proportional grabs from the line and injects them into a constant-pressure receiver, so the receiver fills gradually over the batch with a composite of the whole transfer. The receiver is sized so the batch fills it to the planned level, with the same fill limit as a manual cylinder, and then goes to the laboratory like any floating-piston cylinder.

Common mistakes at the sample point

  • Connecting a cylinder without purging the connection, so air is trapped in the sample.

  • Filling against a precharge that is too low, so the sample flashes as it enters.

  • Opening the inlet valve quickly, causing a sharp pressure drop across it.

  • Overfilling, leaving no room for thermal expansion.

  • Shipping cylinders without checking piston seals, valve caps and labels.

Materials, seals, transport and unloading

  • Stainless steel bodies are standard; NACE MR0175/ISO 15156 materials apply in sour service.

  • Trace sulphur compounds can adsorb on bare metal, so internally treated or coated surfaces are used where those compounds are being measured.

  • Piston and valve seals must suit the product, any methanol or aromatics present and the temperature range; confirm seal materials with the cylinder maker.

  • Cylinders must be rated above the maximum line pressure, with relief devices where the design calls for them.

Sample cylinders are pressure receptacles, so their shipment must meet the transport rules of the region, and labels and test dates should be checked before each trip. In the laboratory, emptying a cylinder by cracking a valve in a fume cupboard is common and avoidable. A sample cylinder unloading station such as the TI-SNE-20 lets technicians depressurise, sub-sample and empty cylinders in a closed set-up, with nitrogen purge and vents routed to extraction or a closed system.

Build the whole sampling chain

We design pressurised hydrocarbon automatic sampling systems for condensate, NGL and LPG and LPG, LNG and hydrogen sampling systems in Vaughan, Canada and build them in Chennai, India, together with receivers, cylinders and the laboratory unloading station. Each system is factory acceptance tested, witnessed by the client or a third-party inspector, with EN 10204 3.1 material certificates. For gas-phase sampling, see our guide to gas sampling stations, and for toxic liquid service, closed-loop sampling for H2S and benzene. Browse our automatic sampling systems, sampling systems and the full oil and gas systems range. Send your product, line pressure and temperature, and analysis requirements to mail@techincresearch.com or call +91 739 749 8656 and we will return a P&ID, GA drawing and priced offer.