A pretreatment trial asks a simple question: which option, a media filter, a UF stage or a different coagulant dose, takes the most fouling load out of the RO feed? The answer has to be a number that can be compared from one option to the next. SDI struggles here; the modified fouling index, MFI-0.45, does not. The FILTON 4750 gives a pilot plant or a trial rig a fixed, repeatable MFI test on the feed line.

Why MFI suits a trial

SDI is not proportional to the particulate load: halving the particles in the water does not halve the SDI. It also moves with pressure, temperature and the filter membrane used, so small differences between test runs blur the comparison.

MFI-0.45 is taken from the cake-filtration part of a constant-pressure test through a 0.45 µm filter: the slope of the plot of time/volume against volume, in s/L². It rises in step with the particulate load and is corrected for temperature and pressure. If one pretreatment option halves the MFI, it has roughly halved the particulate fouling load, which is the comparison a trial needs.

What the FILTON 4750 does

The FILTON 4750 is built on the same platform as our Silton 4750 online SDI analyser:

  1. The operator fits a fresh 47 mm, 0.45 µm disc in the stainless filter holder and starts the test.

  2. The booster pump and regulator hold 2.07 bar (30 psi) at the filter, the ASTM D4189 test condition, even on a low-pressure sample line.

  3. The analyser logs the filtered volume against time throughout the test.

  4. It fits the straight-line part of the curve, shows MFI-0.45 on the screen and sends it on 4–20 mA to the PLC or HMI, with an over-limit alarm and an RO membrane bypass alarm.

Every result is stored, so a trial builds its own record without anyone copying numbers from a stopwatch.

Getting repeatable results

  • Use the same filter type for every test. The membrane's own resistance affects the early part of the curve; keep it constant.

  • Flush the sample line before each test so the water in the holder is the water you mean to measure.

  • Log the operating state of the pretreatment at the time of each test, so MFI can be matched to the setting that produced it.

When to move to the FILTON 4950 Pro

The FILTON 4750 needs an operator to change the disc. When MFI moves from the trial to the running plant and should be measured every hour without anyone at the instrument, the FILTON 4950 Pro changes its own filter from a roll. For SDI instead of MFI, see the SDI Measurement Instruments range.

What to send us

Tell us the water source, the line pressure at the sample point, your power supply and how results should reach your control system. Write to mail@techincresearch.com or call +1 (289) 622-1032.

Specifications are typical and may change. Results depend on the feed water, how the instrument is used and operation in line with the operation manual.