Membrane Test Cells Compared: Active Area, Pressure, Material and Process


Bottom line: for UF and NF coupons in dead-end mode, use a high-pressure stirred cell. For RO and NF under crossflow, the 42 cm² crossflow cell and the 3 × 7 cm SS316 cell are rated to 69 bar (1,000 psi) in SS316. MD, pervaporation, ceramic tubes and spiral-wound elements each have their own cell or housing below.
Stirred (dead-end) cells
Cell | Processes | Membrane | Max pressure | Material | Notes |
|---|---|---|---|---|---|
UF | flat-sheet coupons | up to 10 bar | stainless steel | kit: pressurised reservoir, tubing, fittings, spare O-rings, starter coupons | |
UF, NF | flat-sheet coupons | up to 10 bar | stainless steel | cell body only; magnetic stirring | |
Low-pressure solvent-resistant stirred cell, 25 mm: with stirrer or without | solvent testing | 25 mm | low pressure | solvent resistant | magnetic stirrer optional |
Low-pressure solvent-resistant stirred cell, 47 mm: with stirrer or without | solvent testing | 47 mm | low pressure | solvent resistant | magnetic stirrer optional |
Crossflow and specialist flat-sheet cells
Cell | Processes | Active area | Max pressure | Material | Notes |
|---|---|---|---|---|---|
high pressure in SS316; low-pressure UF and MF in acrylic | 42 cm² | up to 69 bar (1,000 psi) in SS316 | SS316 or acrylic | up to 80 °C (176 °F); spacers 17 to 65 mil; heating jacket option | |
high pressure in SS316; low pressure in acrylic | 140 cm² | up to 69 bar (1,000 psi) | SS316 or acrylic | hydraulic holder; up to 80 °C (176 °F); shim set for channel height | |
RO | 3 × 7 cm | up to 69 bar (1,000 psi) | SS316 | rectangular crossflow channel; separate permeate port | |
RO | 3 × 7 cm | up to 69 bar (1,000 psi) | acetal polymer | metal-free; lighter than steel | |
RO, FO | 4 × 10 cm | up to 69 bar (1,000 psi) | SS316 | dual-purpose cell | |
RO, NF, FO | 3 × 7 cm | up to 69 bar | stainless steel | FO with flow on both sides of the membrane | |
DCMD, AGMD, SGMD | 10 × 15 cm | up to 3 bar | SS316 | separate hot and cold chambers | |
MD | 1.1 m² | – | – | temperature ports at several points; module-scale validation | |
pervaporation | 7 cm diameter | up to 3 bar | SS316 | vacuum-compatible permeate port |
Tubular and element housings
Housing | Processes | Element size | Max pressure | Material | Notes |
|---|---|---|---|---|---|
ceramic membrane tubes | tubes 10 to 25 mm, up to 1.2 m | up to 10 bar (145 psi) | SS316 | up to 120 °C (248 °F), seal dependent; single, multi-tube or jacketed | |
RO, NF | 1812 to 8040 elements | up to 69 bar (1,000 psi) RO; 21 bar (305 psi) NF | SS316 or FRP | side or end ports |
Accessory: the test cell temperature control enclosure holds one or two cells at ambient + 5 to 90 °C (41 to 194 °F), stable to about ±0.5 °C (±0.9 °F).
How to choose
Dead-end or crossflow: stirred cells suit quick UF and NF screening; crossflow cells reproduce the shear of a real module.
Pressure rating must exceed the osmotic pressure of your feed plus the driving pressure you need.
Active area: small cells save membrane and feed; larger cells give more permeate and steadier flux data.
Material: stainless steel for high pressure and solvents; polymer cells for metal-free tests.
Pick a skid that matches the cell's connections and pressure rating.
All cells: membrane test cells. By process: MF, UF, NF, RO, FO, MD and pervaporation test cells.
Before you order
Send us your samples, operating conditions and the standard you work to, and we will confirm the model. Equipment carries our standard warranty and installation, training and service terms: see Service and Support and our Terms of Sale. Send an enquiry or email mail@techincresearch.com.
Data are typical and taken from each product page and datasheet; final design is to the client specification and the order acknowledgement. Check the product page or ask us before you order.