top of page

Membrane Test Cells Compared: Active Area, Pressure, Material and Process

Writer: Tech Inc
Tech Inc
6 hours ago
3 min read
Membrane Test Cells Compared: Active Area, Pressure, Material and Process

Bottom line: for UF and NF coupons in dead-end mode, use a high-pressure stirred cell. For RO and NF under crossflow, the 42 cm² crossflow cell and the 3 × 7 cm SS316 cell are rated to 69 bar (1,000 psi) in SS316. MD, pervaporation, ceramic tubes and spiral-wound elements each have their own cell or housing below.


Stirred (dead-end) cells

Cell

Processes

Membrane

Max pressure

Material

Notes

UF

flat-sheet coupons

up to 10 bar

stainless steel

kit: pressurised reservoir, tubing, fittings, spare O-rings, starter coupons

UF, NF

flat-sheet coupons

up to 10 bar

stainless steel

cell body only; magnetic stirring

Low-pressure solvent-resistant stirred cell, 25 mm: with stirrer or without

solvent testing

25 mm

low pressure

solvent resistant

magnetic stirrer optional

Low-pressure solvent-resistant stirred cell, 47 mm: with stirrer or without

solvent testing

47 mm

low pressure

solvent resistant

magnetic stirrer optional


Crossflow and specialist flat-sheet cells

Cell

Processes

Active area

Max pressure

Material

Notes

high pressure in SS316; low-pressure UF and MF in acrylic

42 cm²

up to 69 bar (1,000 psi) in SS316

SS316 or acrylic

up to 80 °C (176 °F); spacers 17 to 65 mil; heating jacket option

high pressure in SS316; low pressure in acrylic

140 cm²

up to 69 bar (1,000 psi)

SS316 or acrylic

hydraulic holder; up to 80 °C (176 °F); shim set for channel height

RO

3 × 7 cm

up to 69 bar (1,000 psi)

SS316

rectangular crossflow channel; separate permeate port

RO

3 × 7 cm

up to 69 bar (1,000 psi)

acetal polymer

metal-free; lighter than steel

RO, FO

4 × 10 cm

up to 69 bar (1,000 psi)

SS316

dual-purpose cell

RO, NF, FO

3 × 7 cm

up to 69 bar

stainless steel

FO with flow on both sides of the membrane

DCMD, AGMD, SGMD

10 × 15 cm

up to 3 bar

SS316

separate hot and cold chambers

MD

1.1 m²

–

–

temperature ports at several points; module-scale validation

pervaporation

7 cm diameter

up to 3 bar

SS316

vacuum-compatible permeate port


Tubular and element housings

Housing

Processes

Element size

Max pressure

Material

Notes

ceramic membrane tubes

tubes 10 to 25 mm, up to 1.2 m

up to 10 bar (145 psi)

SS316

up to 120 °C (248 °F), seal dependent; single, multi-tube or jacketed

RO, NF

1812 to 8040 elements

up to 69 bar (1,000 psi) RO; 21 bar (305 psi) NF

SS316 or FRP

side or end ports


Accessory: the test cell temperature control enclosure holds one or two cells at ambient + 5 to 90 °C (41 to 194 °F), stable to about ±0.5 °C (±0.9 °F).


How to choose

  • Dead-end or crossflow: stirred cells suit quick UF and NF screening; crossflow cells reproduce the shear of a real module.

  • Pressure rating must exceed the osmotic pressure of your feed plus the driving pressure you need.

  • Active area: small cells save membrane and feed; larger cells give more permeate and steadier flux data.

  • Material: stainless steel for high pressure and solvents; polymer cells for metal-free tests.

  • Pick a skid that matches the cell's connections and pressure rating.


All cells: membrane test cells. By process: MF, UF, NF, RO, FO, MD and pervaporation test cells.


Before you order

Send us your samples, operating conditions and the standard you work to, and we will confirm the model. Equipment carries our standard warranty and installation, training and service terms: see Service and Support and our Terms of Sale. Send an enquiry or email mail@techincresearch.com.


Data are typical and taken from each product page and datasheet; final design is to the client specification and the order acknowledgement. Check the product page or ask us before you order.

Comments


bottom of page