Membrane Filter Chemical Compatibility Chart

At a glance. The wrong membrane can swell, dissolve or shed extractables into a sample. This chart rates ten membrane materials against thirteen common acids, bases and solvent groups at room temperature and short contact. Resistant means suitable; Limited means test first, as swelling or extractables are possible; No means not recommended. Each material links to its syringe filters, membrane disc filters or capsule filters.
Chart 1: PES, PVDF, PTFE, nylon and regenerated cellulose
Chemical | PES | PVDF | PTFE | Nylon | RC |
Dilute acids | Resistant | Resistant | Resistant | Limited | Resistant |
Concentrated acids | Limited | Limited | Resistant | No | No |
Dilute bases | Resistant | Resistant | Resistant | Resistant | Resistant |
Concentrated bases | Limited | No | Resistant | Limited | No |
Alcohols (methanol, ethanol, IPA) | Resistant | Resistant | Resistant | Resistant | Resistant |
Acetonitrile | Limited | Resistant | Resistant | Resistant | Resistant |
Acetone and other ketones | No | Limited | Resistant | Resistant | Resistant |
Esters (ethyl acetate) | No | Limited | Resistant | Resistant | Resistant |
Chlorinated solvents (DCM, chloroform) | No | Limited | Resistant | Limited | Resistant |
Aromatic hydrocarbons (toluene) | No | Resistant | Resistant | Resistant | Resistant |
Aliphatic hydrocarbons (hexane) | Resistant | Resistant | Resistant | Resistant | Resistant |
Tetrahydrofuran (THF) | No | Limited | Resistant | Resistant | Resistant |
DMF, DMSO and NMP | No | No | Resistant | Resistant | Resistant |
Chart 2: MCE, cellulose acetate, polypropylene, glass fibre and track-etched polycarbonate
Chemical | MCE | CA | PP | Glass fibre | PCTE |
Dilute acids | Resistant | Limited | Resistant | Resistant | Resistant |
Concentrated acids | No | No | Limited | Limited | Limited |
Dilute bases | Limited | Limited | Resistant | Resistant | Limited |
Concentrated bases | No | No | Resistant | No | No |
Alcohols (methanol, ethanol, IPA) | Limited | Limited | Resistant | Resistant | Limited |
Acetonitrile | No | No | Resistant | Resistant | Limited |
Acetone and other ketones | No | No | Resistant | Resistant | Limited |
Esters (ethyl acetate) | No | No | Resistant | Resistant | Limited |
Chlorinated solvents (DCM, chloroform) | Limited | Limited | Limited | Resistant | No |
Aromatic hydrocarbons (toluene) | Resistant | Limited | Limited | Resistant | Limited |
Aliphatic hydrocarbons (hexane) | Resistant | Resistant | Resistant | Resistant | Resistant |
Tetrahydrofuran (THF) | No | No | Limited | Resistant | No |
DMF, DMSO and NMP | No | No | Resistant | Resistant | No |
How to read the chart
Find the strongest chemical in the sample, not the main one. A mostly aqueous sample with 30% acetonitrile is an acetonitrile sample for the filter.
Pick a membrane rated Resistant for every component. If only Limited ratings are left, run a blank: filter the solvent alone and check the filtrate by your method.
Check the housing. Syringe and capsule filters have polypropylene housings, which limit use with hot chlorinated and aromatic solvents even when the membrane is PTFE.
Check wetting. Hydrophobic PTFE and PVDF do not wet with water. Pre-wet with methanol or isopropanol, then flush with water, or choose a hydrophilic version.
Check binding. For proteins and drugs at low concentration, PES, PVDF, regenerated cellulose and cellulose acetate bind least; nylon and MCE bind most.
Which membrane for which job
Aqueous and biological samples: PES syringe filters for speed and low binding; cellulose acetate when protein recovery matters most.
Reversed-phase HPLC samples: PVDF, regenerated cellulose or nylon for non-protein samples.
Aggressive solvents, strong acids and bases: PTFE syringe filters, PTFE discs and PTFE capsules.
Trace analysis with acids and bases: all-polypropylene syringe filters.
Microbiology, SDI and particle counts: MCE membrane discs and track-etched polycarbonate.
Dirty or viscous samples: glass-fibre prefilter syringe filters.
Temperature
Ratings fall as temperature rises. Most membranes are rated to about 40 to 60 °C in contact with solvents, and polypropylene housings soften well before PTFE does. For hot solvents, use PTFE in a stainless steel filter holder.
Frequently asked questions
Is nylon compatible with acetonitrile? Yes. Nylon resists acetonitrile and most HPLC solvents, but not strong acids.
Can PES be used with methanol? Yes for alcohols at room temperature; avoid ketones, esters, chlorinated solvents and DMSO.
Which filter is safe for DMSO or DMF? PTFE, nylon and regenerated cellulose. PES, PVDF, MCE and cellulose acetate are not recommended.
Why test even when the chart says Resistant? Ratings cover the membrane in short contact. Long soaks, heat, mixtures and trace-level analysis can still show effects, so a blank run is the safest check.
Send your sample and solvent to mail@techincresearch.com and our engineers will recommend the membrane, pore size and format.