top of page

Laboratory Gas Generators Compared: Nitrogen, Hydrogen, Zero Air and Liquid Nitrogen

Writer: Tech Inc
Tech Inc
8 hours ago
2 min read
Laboratory Gas Generators Compared: Nitrogen, Hydrogen, Zero Air and Liquid Nitrogen

Bottom line: one GC needs hydrogen and zero air, so the GC Gas Station or the three-in-one generator replaces several cylinders. LC-MS needs the nitrogen generator, and a hydrogenation reactor needs the high-pressure hydrogen generator with its booster.


Side by side

Generator

Gases

Output

Outlet pressure

Technology

Utilities

nitrogen

up to 35 L/min

to the LC-MS specification

hollow-fibre membrane modules; built-in oil-free compressor

mains power

hydrogen

sized to the GC duty

up to 7 bar (102 psi)

PEM, no liquid caustic electrolyte

deionised water, mains power

hydrogen and zero air

sized to the GC and detector duty

to the GC specification

PEM hydrogen; catalytic zero air

room air, deionised water, mains power

hydrogen, nitrogen, zero air

hydrogen about 250 or 500 mL/min by model

three regulated outlets with gauges

PEM; membrane or PSA nitrogen; catalytic zero air

single-phase supply

hydrogen

about 0.5 to 1 L/min

generator about 16 bar (232 psi); booster about 100 bar (1,450 psi)

PEM with booster and buffer vessel

–

liquid nitrogen

about 5 to 200 L/day

–

cryocooler; nitrogen about 99.5% or better

single- or three-phase


How to choose

  • Add up the flow of every instrument the generator will serve, with a margin.

  • Check purity and pressure against each instrument maker's requirement.

  • Hydrogen generators hold very little gas and shut down on a leak or pressure fault; all three hydrogen units have leak detection and shut-down.

  • PEM units need deionised water; nitrogen units need room air or a compressed-air supply.



Before you order

Send us your samples, operating conditions and the standard you work to, and we will confirm the model. Equipment carries our standard warranty and installation, training and service terms: see Service and Support and our Terms of Sale. Send an enquiry or email mail@techincresearch.com.


Data are typical and taken from each product page and datasheet; final design is to the client specification and the order acknowledgement. Check the product page or ask us before you order.