Laboratory Gas Generators Compared: Nitrogen, Hydrogen, Zero Air and Liquid Nitrogen


Bottom line: one GC needs hydrogen and zero air, so the GC Gas Station or the three-in-one generator replaces several cylinders. LC-MS needs the nitrogen generator, and a hydrogenation reactor needs the high-pressure hydrogen generator with its booster.
Side by side
Generator | Gases | Output | Outlet pressure | Technology | Utilities |
|---|---|---|---|---|---|
nitrogen | up to 35 L/min | to the LC-MS specification | hollow-fibre membrane modules; built-in oil-free compressor | mains power | |
hydrogen | sized to the GC duty | up to 7 bar (102 psi) | PEM, no liquid caustic electrolyte | deionised water, mains power | |
hydrogen and zero air | sized to the GC and detector duty | to the GC specification | PEM hydrogen; catalytic zero air | room air, deionised water, mains power | |
hydrogen, nitrogen, zero air | hydrogen about 250 or 500 mL/min by model | three regulated outlets with gauges | PEM; membrane or PSA nitrogen; catalytic zero air | single-phase supply | |
hydrogen | about 0.5 to 1 L/min | generator about 16 bar (232 psi); booster about 100 bar (1,450 psi) | PEM with booster and buffer vessel | – | |
liquid nitrogen | about 5 to 200 L/day | – | cryocooler; nitrogen about 99.5% or better | single- or three-phase |
How to choose
Add up the flow of every instrument the generator will serve, with a margin.
Check purity and pressure against each instrument maker's requirement.
Hydrogen generators hold very little gas and shut down on a leak or pressure fault; all three hydrogen units have leak detection and shut-down.
PEM units need deionised water; nitrogen units need room air or a compressed-air supply.
All models: laboratory gas generator range.
Before you order
Send us your samples, operating conditions and the standard you work to, and we will confirm the model. Equipment carries our standard warranty and installation, training and service terms: see Service and Support and our Terms of Sale. Send an enquiry or email mail@techincresearch.com.
Data are typical and taken from each product page and datasheet; final design is to the client specification and the order acknowledgement. Check the product page or ask us before you order.