Gurley Test Explained: Gurley Seconds, Air Permeability and How to Run It

Writer: Tech Inc
Tech Inc
1 day ago
4 min read

At a glance. A Gurley test times how long 100 mL of air takes to pass through about 6.45 cm² (1 in²) of a sheet under a pressure difference of about 1.22 kPa. The result, in Gurley seconds, is an air resistance: the higher the number, the less permeable the material. It is defined in TAPPI T 460 and ISO 5636-5 for paper and is the everyday porosity check for battery separators, membranes, nonwovens and filter media. It is quick and simple, but it is a comparative number, not a pore size: for pore size and bubble point, pair it with a capillary flow porometer.

What does the Gurley test measure?

The Gurley test measures air resistance: the time for a known volume of air to flow through a known area of material under a known, small pressure. Air finds its way through every connected path, so the result reflects porosity, pore size, tortuosity and thickness together.

  • Low Gurley seconds mean an open, permeable sheet (tissue, coarse filter media, highly porous separators).

  • High Gurley seconds mean a tight, low-permeability sheet (greaseproof papers, dense coatings, low-porosity separators).

Because it bundles several properties into one number, Gurley is best used to compare like with like: one batch of separator against the last, one casting run against another, one supplier against a specification.

Gurley test standards and conditions

Standard

Material

Air volume

Test area

Pressure difference

Result

TAPPI T 460

Paper

100 mL (typical)

About 6.45 cm² (1 in²)

1.22 kPa

Seconds per 100 mL (recommended range 5 to 1800 s)

ISO 5636-5

Paper and board

100 mL (typical)

About 6.4 cm²

About 1.23 kPa

Seconds, convertible to air permeance

Separator specifications

Battery separators

Usually 100 mL

Per Gurley densometer

Per Gurley densometer

Seconds per 100 mL, often normalised to thickness

TAPPI notes that very rough sheets that cannot be clamped without leakage are unsuitable, and that very impermeable materials take impractically long; for those, TAPPI T 536 (higher pressure) and TAPPI T 547 (up to about 9.85 kPa, several test areas) are used.

Gurley seconds, air permeance and porosity

Gurley seconds are inversely related to air permeance. For a given test area and pressure, halving the Gurley time means roughly twice the air flow. Many laboratories convert results to air permeance (for example µm/(Pa·s)) so that tests at different areas or pressures can be compared.

Gurley does not give porosity or pore size directly. Two sheets with the same porosity can have very different Gurley values if one has fewer, smaller or more tortuous pores, or is thicker. To relate structure to Gurley:

  • measure thickness, and report Gurley per unit thickness when comparing grades;

  • measure porosity by weight and density, or by porosity instrument;

  • measure bubble point and mean flow pore size with a capillary flow porometer.

Where the Gurley test is used

  • Lithium-ion battery separators. Gurley is the standard quick check of separator permeability and uniformity in production and incoming inspection. Too high a value restricts ion transport; too low can indicate defects or poor shutdown behaviour.

  • Membranes. A fast screen of flat-sheet and electrospun nanofibre membranes between casting runs, before slower water flux and pore size tests.

  • Filter media and nonwovens. Air resistance of meltblown, spunbond and glass-fibre media.

  • Paper and board. The original application: printing, packaging and greaseproof grades.

How to run a reliable Gurley test

  1. Condition the samples. Paper is conditioned to TAPPI or ISO atmospheres; membranes and separators should be dry and free of wetting liquid, which blocks pores.

  2. Cut clean specimens larger than the clamp, free of creases, pinholes and edge damage.

  3. Check for leaks with an impermeable reference sheet; any flow is leakage, not permeability.

  4. Clamp evenly. Uneven clamping on rough or thick samples lets air bypass the test area and gives falsely low values.

  5. Choose the volume so that the timing is neither too short to time accurately nor impractically long; report the volume used.

  6. Test several locations across the width and length of the roll, and report the mean and spread.

  7. Calibrate the instrument with reference orifices or standards at the interval your quality system requires.

Gurley tester or porometer: which do you need?

Question

Gurley tester

Capillary flow porometer

Fast pass/fail permeability check

Yes

Possible, slower

Bubble point (largest pore)

No

Yes

Mean flow pore size and distribution

No

Yes

Gas permeability curve vs pressure

Single point

Full curve

Typical use

Production QC, incoming inspection

R&D, specification, failure analysis

Most separator and membrane laboratories use both: Gurley on every roll, porometry on qualification samples and whenever Gurley drifts.

Frequently asked questions

What is a Gurley number?

The time in seconds for a set volume of air, usually 100 mL, to pass through a set area of sheet at a set pressure, as defined by the Gurley method.

Is a higher Gurley value better?

Neither: it depends on the product. A higher value means lower air permeability. Separators and membranes have target windows set by the cell or process design.

Does the Gurley test measure pore size?

No. It measures air resistance, which depends on pore size, porosity, tortuosity and thickness together. Use a capillary flow porometer for pore size.

Why do two labs get different Gurley results on the same sample?

Usually because of leakage at the clamp, different test areas or volumes, sample conditioning, or sampling different parts of the roll.