Porosity Instruments: How Porosity Testers Work and How to Choose One

Updated: 23 hours ago
At a glance. "Porosity instrument" covers five families of machine, and they do not measure the same thing. A capillary flow porometer measures the size of pores that run right through a material (bubble point, mean flow pore size, pore size distribution) and is the standard tool for membranes and filter media. A mercury intrusion porosimeter measures total open porosity and pore volume distribution in rigid solids. A gas pycnometer gives skeletal density, from which porosity is calculated. A gas adsorption analyser measures surface area and nanometre-scale pores. An Archimedes (water immersion) tester gives bulk density and open porosity of ceramics and sintered metals. Choose by the property you need, the pore size range and whether the sample can be wetted, pressurised or destroyed.
What does a porosity instrument measure?
Porosity is the fraction of a material's volume that is empty space. Most buyers actually need one of four different numbers:
Total or open porosity (%): how much of the volume is pore space that a fluid can reach.
Through-pore size: the size of the narrowest constriction in pores that connect one face to the other. This is what controls filtration, rejection and flow.
Pore size distribution: how pore volume or flow is spread across pore sizes.
Density: bulk (envelope) and skeletal density, from which porosity is calculated.
A membrane maker usually cares about through-pores. A ceramic or powder-metallurgy plant usually cares about open porosity and density. Buying the wrong class of porosity tester is the most common and expensive mistake in this category.
The five types of porosity tester compared
Instrument | What it measures | Typical pore range | Key standards | Best for |
Capillary flow porometer | Bubble point, mean flow pore size, through-pore size distribution, gas permeability | About 0.01 to 500 µm | ASTM F316, ISO 2942, ISO 4003 | Membranes, filter media, nonwovens, battery separators, sintered filters |
Mercury intrusion porosimeter | Total open porosity, pore volume distribution, bulk and skeletal density | About 3 nm to several hundred µm | ISO 15901-1, ASTM D4284, ASTM D4404 | Catalysts, ceramics, rock and soil, powders, rigid solids |
Gas pycnometer | Skeletal (true) density, closed porosity when combined with bulk density | Not a pore sizer | ISO 12154, ASTM B923, ASTM C604 | Metal powders, ceramics, pharmaceuticals, foams |
Gas adsorption (BET) analyser | Specific surface area, micropores and mesopores | About 0.35 to 100 nm | ISO 9277, ISO 15901-2, ISO 15901-3 | Adsorbents, catalysts, activated carbon, MOFs |
Archimedes density tester | Bulk density, apparent (open) porosity, water absorption | Not a pore sizer | ASTM B962, ASTM B328, ASTM C20, ASTM C373 | Sintered metal parts, refractories, fired ceramics |
Ranges are typical for commercial instruments and depend on the model, the wetting liquid or gas, and the maximum pressure.
Capillary flow porometer: the porosity tester for membranes and filters
A capillary flow porometer fills the pores of the sample with a wetting liquid of low surface tension, then raises gas pressure across it. The largest pore empties first; that pressure is the bubble point. As pressure rises, smaller pores open and gas flow increases until the sample behaves like the dry sample. Comparing the wet and dry flow curves gives the mean flow pore size and the full through-pore size distribution, using the Young–Laplace relation between pressure and pore diameter.
Because only pores that connect both faces carry gas, a porometer reports exactly the pores that govern filtration. The test is fast (minutes per sample), uses no mercury and leaves the sample intact. It is the method behind ASTM F316 for membrane filters, ISO 2942 for hydraulic filter elements and ISO 4003 for permeable sintered metals.
Tech Inc. builds the AL500 EFM porosity instrument, the AL1000 and AL1000 EFM, and the skid-mounted AL500 pore size and porosity system for membrane, filter and separator laboratories. The full range is on the porometers page.
Mercury intrusion porosimeter
A mercury porosimeter forces non-wetting mercury into an evacuated sample at rising pressure, up to about 400 MPa on high-pressure units. The volume intruded at each pressure gives pore volume against pore entry size, plus total porosity and density. It reaches very small pores and suits rigid solids, but it measures pore entrances (so ink-bottle pores are undersized), compresses soft polymers, destroys the sample and requires mercury handling and disposal. For a side-by-side view, see capillary flow porometry vs mercury intrusion porosimetry and what a mercury intrusion porosimeter is used for.
Gas pycnometer
A helium pycnometer measures the volume of the solid itself by gas displacement. With the bulk (envelope) volume from geometry or another method, porosity follows directly:
Porosity (%) = (1 − bulk density ÷ skeletal density) × 100
It is quick and precise, but it does not size pores. It is the usual metal porosity measurement instrument for powders and the reference for closed porosity in foams and ceramics.
Gas adsorption (BET) analyser
Nitrogen, argon or CO₂ adsorption at cryogenic temperature gives specific surface area by the BET method and the size distribution of micropores and mesopores. It is essential for catalysts, adsorbents and activated carbon, but the pores it sees are mostly far smaller than those that control membrane filtration, and it says little about through-pores.
Archimedes density and porosity tester
The sample is weighed dry, then saturated with water (often by boiling or vacuum), weighed suspended in water and weighed saturated in air. Bulk density, apparent porosity and water absorption follow from the three masses. It is simple and inexpensive, and it is the routine porosity measurement for sintered metal parts (ASTM B962, ASTM B328), refractory brick (ASTM C20) and fired ceramic tiles (ASTM C373). It gives no pore size information.
How to choose a porosity testing machine
Name the property. Through-pore size and bubble point: porometer. Total porosity and pore volume: mercury porosimeter or Archimedes. Density: pycnometer. Surface area and nanopores: gas adsorption.
Check the pore size range. Match the instrument's range and maximum pressure to the smallest pore you must resolve, with margin.
Consider the sample. Thin, flexible or compressible membranes and nonwovens suit porometry. Rigid ceramics, rock and catalysts suit mercury intrusion. Parts too large for a cell suit Archimedes.
Decide if the test may be destructive. Porometry and pycnometry keep the sample; mercury intrusion does not.
Follow your customer's standard. If a specification calls for ASTM F316 or ISO 2942, a bubble point and mean flow pore test is required, not a mercury curve.
Plan for throughput. Quality-control labs value automatic wet-up, dry-up and reporting; research labs value wide pressure range and multiple wetting liquids.
Count the running costs. Wetting liquids and gas for porometers; mercury handling, waste and safety for porosimeters; helium for pycnometers.
What does a porosimeter or porometer cost?
Price is driven by the measurement principle, the maximum pressure (and so the smallest pore), the number of sample holders, automation and software, and whether the unit is benchtop or skid-mounted. Mercury porosimeters carry extra costs for mercury safety and waste. For a quotation on a Tech Inc. capillary flow porometer, send your sample type, thickness, expected pore size and the test standard you follow, to mail@techincresearch.com.
Frequently asked questions
What instrument is used to measure porosity?
It depends on the material. Membranes and filters are tested with a capillary flow porometer; rigid solids with a mercury intrusion porosimeter; powders and ceramics with a gas pycnometer or the Archimedes method.
What is the difference between a porometer and a porosimeter?
A porometer measures through-pores by gas flow through a wetted sample. A porosimeter (usually mercury intrusion) measures all pores a liquid can enter, including dead-end pores, and reports pore volume.
How is porosity of metal measured?
Sintered metal parts are usually tested by the Archimedes method (ASTM B962, ASTM B328). Porous sintered metal filters are tested for bubble point and pore size by porometry (ISO 4003). Powders are measured by gas pycnometry (ASTM B923).
Can one porosity tester do everything?
No. Each method answers a different question. Many laboratories pair a capillary flow porometer for through-pores with a pycnometer or mercury porosimeter for total porosity.