When water passes through a membrane and salt does not, salt builds up in a thin layer at the membrane surface. That layer, concentration polarisation, makes a membrane look worse than it is and makes test results depend on the test cell as much as the membrane.

The film model

The surface concentration Cm relates to the bulk concentration Cb by:

Cm ÷ Cb = exp(Jv ÷ k)

  • Jv: permeate flux (as a velocity, m/s).

  • k: mass transfer coefficient (m/s), which rises with crossflow velocity and good feed-channel design.

A 2018 review by Belfort writes the same relation as the concentration modulus and notes that k varies with crossflow velocity.

Observed and intrinsic rejection

  • Observed rejection, Ro = 1 − Cp ÷ Cb: what you measure from the tank and the permeate.

  • Intrinsic rejection, Ri = 1 − Cp ÷ Cm: what the membrane itself does.

Worked example (illustrative numbers)

  • Flux 20 LMH = 20 ÷ 3,600 ÷ 1,000 = 5.56 × 10⁻⁶ m/s.

  • Assumed k = 2.0 × 10⁻⁵ m/s.

  • Polarisation factor = exp(5.56 × 10⁻⁶ ÷ 2.0 × 10⁻⁵) = exp(0.278) = 1.32.

  • Cb = 2,000 mg/L, so Cm = 2,640 mg/L.

  • Cp = 40 mg/L: Ro = 98.0%, Ri = 1 − 40 ÷ 2,640 = 98.5%.

What it means for testing

  • Compare membranes at the same flux and crossflow velocity, or the comparison measures the test conditions, not the membranes.

  • Higher flux, more polarisation. Raising pressure raises flux and Cm together, so rejection can look worse at high flux.

  • Cell design matters. Feed spacers and channel height set k. A large crossflow cell with a defined channel gives repeatable conditions.

  • Polarisation also drives fouling and scaling, because foulants and sparingly soluble salts concentrate at the surface first.

Write to mail@techincresearch.com or call +1 (289) 622-1032.

Worked examples use illustrative numbers. Equipment specifications are typical and may change; results depend on the membranes and conditions tested, how the equipment is used and operation in line with the operation manual.