CO2 to Fuels in the Laboratory: Designing an RWGS and Fischer-Tropsch Catalyst Test Unit

At a glance. Making liquid fuel from captured carbon dioxide is a two-step route. Reverse water gas shift (RWGS) turns carbon dioxide and hydrogen into carbon monoxide and water; Fischer-Tropsch synthesis turns the resulting synthesis gas into hydrocarbons. A laboratory unit that runs both reactions, alone or in series, lets one team screen catalysts for each step and then study the coupled route. Most of what decides the quality of the data is not the reactor itself but the feed, the heated product path, the analysis and the safety layer around it.
The two reactions
Reverse water gas shift: CO2 + H2 → CO + H2O. Endothermic and favoured by high temperature, so the reactor runs up to about 900 °C and is made of a nickel alloy such as alloy 625.
Fischer-Tropsch synthesis: nCO + (2n+1)H2 → CnH2n+2 + nH2O. Exothermic, run at lower temperature (up to about 600 °C in this unit), producing gases, liquids and wax.
In series operation the water formed in the first reactor is knocked out before the gas reaches the Fischer-Tropsch catalyst, because water affects that catalyst and the product distribution.
Choosing the pressure
RWGS is mole-neutral: two molecules in, two molecules out. Pressure therefore has no effect on its equilibrium conversion. The operating pressure is set by the Fischer-Tropsch duty and by the analysis train, which is why 40 barg is a sensible figure rather than an arbitrarily higher one. Fewer parts run at high pressure, and the safety case is simpler.
Bringing carbon dioxide to pressure
Carbon dioxide is drawn as liquid from the cylinder, so its supply pressure is capped by its vapour pressure, roughly 51 barg at 15 °C and lower in a cool room. A two-stage oil-free booster on the shared feed header closes the gap, so every gas service reaches the operating pressure, not only the permanent gases.
One manifold for both reactors
Six metered gas services on one shared manifold feed both reactors. That removes duplicated mass flow controllers, leaves one calibration set to maintain, and keeps the gas blend identical whichever reactor is on line, so results from the two reactors can be compared directly.
No cold spot in the product path
Every line from the evaporator to the chromatograph inlet is trace heated, typically at 200 °C, and the wax trap is held above the wax melting point. A single unheated fitting condenses product, and the analysis no longer represents what left the reactor. The heated envelope is a design requirement, not an accessory.
Seeing what the catalyst is doing
Three independently controlled furnace zones per reactor flatten the temperature profile along the bed. A multipoint thermowell inside the bed locates the hot spot, and differential pressure across the bed gives early warning of coking in the RWGS reactor and of wax hold-up in the Fischer-Tropsch reactor.
Analysis
A heated two-position sampling valve sends either reactor outlet to one gas chromatograph with a thermal conductivity detector and two flame ionisation detectors, one with a methaniser for CO and CO2. The stream that is not being analysed continues to the vent without interruption.
A separate safety layer
Each reactor has its own control system and its own hard-wired safety system. The two layers share no element, so a fault in one cannot disable the other. Trips latch and need a manual reset once the cause has cleared, and every final element fails safe on loss of power or signal. Hydrogen and carbon monoxide detectors inside the unit enclosure complete the picture.
Preparing the laboratory
A unit of this size has a footprint of about 4.4 m² and needs a room of at least 20 m². It puts 6 to 9 kW of heat into the room continuously, about 2 to 2.5 tons of refrigeration, so the room's cooling has to be planned before the equipment arrives, along with a 400 V three-phase supply, gas cylinders, demineralised water and a vent routed outside the building.
Specification checklist
Reactions and operating modes: RWGS, Fischer-Tropsch, or both in series
Maximum temperature and pressure for each reactor, and reactor materials
Gas services, ranges and the water feed
Catalyst bed size and particle size range
Analysis: detectors, sampling points and sample conditioning
Safety functions, gas detection and the laboratory's utilities
Where Tech Inc. fits
Tech Inc. engineers and supplies the Integrated RWGS, GTL and Upgrading Reactor Unit, and issues the process flow diagram and the full piping and instrumentation diagram set with the enquiry. Related equipment includes fixed-bed catalyst test units, hydroprocessing pilot plants and high-pressure laboratory hydrogen generators. Send your programme to mail@techincresearch.com.