Application Note: Making Calibration Gases for CEMS and Process Analysers

At a glance. Continuous emission monitoring systems measure SO2, NOx, CO, CO2, O2 and other gases at stacks, and process analysers watch pipelines and units. Regulators and quality schemes require regular checks with zero and span gases of known composition. Specialty-gas makers, analyser companies and large plant laboratories can prepare these mixtures themselves by weight.
Why weight
The composition of a gas mixture can be known very accurately from the masses of each component put into the cylinder. ISO 6142-1 describes this gravimetric method; the mixture is then verified against reference standards, for example by the comparison method of ISO 6143.
Preparing a mixture
Cylinder preparation. The cylinder is heated and evacuated to remove air and moisture, and reactive gases need passivated cylinders.
Empty weight. The cylinder is weighed on a precision balance or mass comparator.
Component additions. Each component, the minor ones first, is added through a clean, leak-tight manifold, and the cylinder is weighed after each addition.
Homogenising. The cylinder is rolled or heated so the gases mix fully.
Calculation and verification. The software computes the composition and uncertainty from the masses and the purity of the parent gases; the laboratory verifies the mixture before certifying it.
Very low concentrations are made in two steps: a pre-mixture first, then a dilution of it.
What the laboratory needs
A blending manifold in electropolished stainless steel with metal seals, a vacuum system, a balance matched to the cylinder size, gas cabinets for toxic components, a stable room and software that records every weighing.