The Tech Inc. Helium Recovery and Purification Skid collects helium that would otherwise be vented from laboratory instruments, magnets, cryostats and leak-test stations, cleans it and stores it for reuse. A gas bag or low-pressure buffer takes the returns, a compressor raises the pressure, and a purifier removes air, moisture and oil before the helium is filled into cylinders or a storage bank.
At a glance: about 5 to 100 Nm³/h · purity typically 99.99% or better · stored at about 150 to 300 bar · adsorption, membrane or cryogenic purification · online purity analysis
Main components
- Recovery piping, flow meters and a gas bag or buffer vessel
- Oil-free or oil-lubricated compressor with after-cooler
- Purifier: adsorption, membrane or cryogenic, to the purity needed
- Purity analyser for helium, oxygen and moisture
- High-pressure filling manifold and storage bank
- PLC with alarms and recovery records
Applications: university and research laboratories, NMR and MRI facilities, cryostat users, helium leak-test stations and industrial users of helium.
Send your process data and specification to mail@techincresearch.com or call +1 (289) 622-1032 for a P&ID, GA drawing and quotation.