Pressure Decay Leak Testing: How It Works and How to Set Limits

At a glance. The part is filled with air to a test pressure, isolated, allowed to stabilise and then watched for a fall in pressure. The leak rate is Q = V·ΔP ÷ (t·Patm), where V is the internal volume. The Tech Inc. Automated Pressure Decay Leak Tester runs the cycle in seconds and logs every result.
The test cycle
Phase | Purpose |
Fill | bring the part to test pressure |
Stabilise | let temperature and stretch settle |
Test | measure pressure decay |
Vent | release pressure safely |
Setting limits
Define the leak rate the product can accept.
Measure the part volume, including fixture volume.
Choose test time long enough to resolve that leak.
Validate with a calibrated leak standard.
When to use other methods
Differential pressure for smaller leaks, comparing against a sealed reference.
Mass flow for larger leaks or big volumes.
Helium for very small leaks.
Frequently asked questions
Why do results drift? Usually temperature: warm parts from upstream processes cool during the test. Longer stabilisation helps.
Can it test filters? For filter integrity, use our integrity tester.