top of page

Capillary Pressure by Porous Plate: The Reference Method

Writer: Tech Inc
Tech Inc
6 hours ago
1 min read

At a glance. Brine-saturated cores sit on a water-wet ceramic plate in a pressure cell. Gas or oil pressure is raised in steps; at each step brine leaves the cores through the plate until equilibrium, and the volume gives saturation. The Tech Inc. Capillary Pressure Porous Plate System tests many cores at once.

Method comparison

Method

Speed

Fluids

Notes

Porous plate

weeks

gas–brine, oil–brine

true equilibrium; reference method

Centrifuge

days

gas–brine, oil–brine

needs data correction

Mercury injection

hours

mercury–vacuum

fast; converted to reservoir fluids

Running a good test

  • Use a plate with entry pressure above the highest step.

  • Ensure good capillary contact (paper or diatomaceous layer).

  • Wait for equilibrium at each step: no more brine collected.

  • Weigh cores at the end to check saturations.

Uses

  • Irreducible water saturation.

  • Saturation-height functions.

  • Preparing cores at known saturation for resistivity index tests.

Frequently asked questions

How many cores per cell? Up to 12 in the large cell.

Why so slow? Flow through the low-permeability plate and the core takes time to reach equilibrium.

Comments