BJH Pore Size and Multi-Station BET: Getting More Isotherms per Day

At a glance. The desorption branch of a nitrogen isotherm shows pores emptying as pressure falls. The Kelvin equation links each pressure to a pore radius, and the BJH method turns the volume released into a pore size distribution from about 2 to 50 nm and beyond. The Tech Inc. Multi-Station BET Surface Area and Pore Size Analyser runs up to three such isotherms at once.
Reading the isotherm
Type I: micropores fill at very low pressure (zeolites, activated carbons).
Type IV with hysteresis: mesopores; the loop shape hints at pore geometry (cylinders, ink-bottles, slits).
Type II: non-porous or macroporous solids.
BJH, t-plot and DFT
Method | Pore range | Best for |
BJH | about 2 to 50 nm and above | mesoporous oxides, catalysts, silicas |
t-plot | micropore volume | zeolites, carbons |
DFT / NLDFT | about 0.4 to 50 nm | combined micro- and mesopores |
Why parallel stations
A full isotherm with 40 to 80 points can take many hours. Independent transducers on each port let three samples equilibrate side by side, so a laboratory triples output without shortening equilibration times.
Specifying an analyser
Number of analysis ports and degas ports.
Micropore option (turbomolecular pump, low-pressure transducer).
Adsorbates: nitrogen, argon, krypton, CO2.
Reporting: BET, BJH, t-plot, DFT and ISO 9277 checks.
Frequently asked questions
Adsorption or desorption branch for BJH? Report both; the desorption branch can be distorted by pore blocking, which the adsorption branch avoids.
Do we need argon? For accurate micropore analysis at 87 K, yes.