The Dean-Stark Method for Core Water and Oil Saturation

At a glance. The core sits above boiling toluene. Toluene vapour carries the core's water up to a condenser; the water collects in a graduated trap while the toluene returns and dissolves the oil. Water volume gives water saturation; the weight loss minus the water gives the oil. The Tech Inc. Dean-Stark Extraction Apparatus comes in three- and six-station frames.
Calculation
Water volume Vw from the trap.
Oil mass = initial weight − dry clean weight − water mass.
Oil volume Vo = oil mass ÷ oil density.
Sw = Vw ÷ pore volume; So = Vo ÷ pore volume.
Good practice
Weigh the core immediately after unwrapping to avoid evaporation losses.
Run until the water level in the trap stops rising.
Continue extraction until the solvent around the core is clear.
Dry the core at a controlled temperature before porosity tests.
Frequently asked questions
Is toluene the only solvent? It is common; others are used for specific oils, following the laboratory's procedure and safety rules.
What comes next? Porosity and grain density on the clean, dry core, for example with our gas pycnometer.
Contact mail@techincresearch.com.