Routine Core Analysis: Automating Porosity and Permeability

At a glance. Helium porosity uses Boyle's law expansion to measure pore volume; gas permeability comes from flow or pressure fall-off through the confined plug, corrected for gas slippage (Klinkenberg). The Tech Inc. Automated Multi-Sample Core Porosimeter and Permeameter runs both on twelve plugs without an operator.
Porosity by helium expansion
Helium at a known pressure expands into the confined plug's pore space. The pressure drop gives pore volume; with bulk volume from calipers or a separate measurement, porosity follows. API RP 40 describes the method.
Permeability by pressure fall-off
Gas pressure upstream of the plug falls as gas flows through. Fitting the fall-off curve gives permeability over a wide range in one test, and several pressures give the Klinkenberg correction.
Why automate
Manual | Automated |
Operator loads every plug | 12-plug carousel |
Timing varies | identical procedure every plug |
Transcription errors | direct data export |
Frequently asked questions
What net stress should be used? Report results at a stated confining pressure; many labs use a standard low stress for screening and reservoir stress for key plugs.
Which plugs suit pulse decay instead? Below about 0.01 mD, use our pulse decay permeameter.