Chemisorption, TPR and TPD: Measuring Catalyst Active Sites

Writer: Tech Inc
Tech Inc
6 hours ago
1 min read

At a glance. A chemisorption analyser passes reactive gas over a catalyst in a small quartz reactor while a furnace ramps the temperature, and a thermal conductivity detector measures what the sample takes up or releases. The Tech Inc. Chemisorption Analyser runs pulse chemisorption, TPR, TPD and TPO on one instrument.

The four techniques

Technique

Gas

What it gives

Pulse chemisorption

H2, CO or O2 pulses

metal dispersion, active metal area, crystallite size

TPR

H2 in argon

reducibility, metal–support interaction

TPD

NH3 or CO2, then heating

acid or base site density and strength

TPO

O2 in helium

coke amount and type on spent catalysts

How a run works

  1. Pre-treat the sample (dry, reduce or oxidise) in the reactor.

  2. Switch to the probe gas.

  3. Ramp the temperature or inject calibrated pulses.

  4. The TCD signal, calibrated against known volumes, gives the amount of gas.

What to specify

  • Furnace range and ramp rates; sub-ambient start for weak sites.

  • Number of mass flow controllers and gas list.

  • Cold trap or mass spectrometer to remove or identify water and products.

  • Single-point BET option for total area on the same sample.

Frequently asked questions

Why not just use BET? BET counts all surface; chemisorption counts only the metal or acid sites that do the chemistry.

Is a mass spectrometer needed? Not for routine work; it helps when several species desorb together.

Pair it with our fixed-bed catalyst test unit for activity tests. Contact mail@techincresearch.com.