Cement Thickening Time and the HPHT Consistometer

At a glance. An HPHT consistometer turns the slurry in a rotating cup inside a pressure vessel while following the well's pressure and temperature schedule. A paddle measures consistency in Bearden units (Bc); thickening time is the time to reach the chosen limit, often 70 or 100 Bc. The Tech Inc. HPHT Cement Consistometer runs API RP 10B-2 schedules automatically.
What the test shows
Thickening time against the planned pumping time plus a safety margin.
Right-angle set: a sharp rise in consistency, preferred for gas-tight cement.
Gelation or bumps that can signal additive problems.
Running good tests
Mix the slurry to the standard procedure.
Choose a schedule matching bottom-hole circulating temperature and pressure.
Calibrate the potentiometer or sensor regularly.
Check temperature calibration of the slurry thermocouple.
Frequently asked questions
What limits apply? Many operators test to 2 to 6 hours and up to about 1400 bar and 260 °C for deep wells.
Atmospheric consistometers? They condition slurries for other tests but do not reproduce downhole pressure.
Contact mail@techincresearch.com.





Comments