The Tech Inc. Time-Lag Gas Permeation Apparatus for CO2 Membranes measures how fast CO2 and other gases pass through a new membrane film, and why. The film is sealed in a cell, both sides are evacuated, and gas is admitted to the feed side at a set pressure. The pressure rise in a fixed downstream volume is recorded against time: its steady slope gives the permeability, and the intercept of the steady line on the time axis, the time lag, gives the diffusivity. Solubility follows from the two. This is the constant-volume, variable-pressure method used to characterise polymer, mixed-matrix and other CO2-selective films.
At a glance: Constant-volume, variable-pressure method · permeability, diffusivity and solubility · ASTM D1434 · ISO 15105-1 · automatic sequences
Main components
- Flat-sheet permeation cell for film coupons, with sealing for thin and brittle films
- Thermostatted oven for the cell and the downstream volume
- Calibrated downstream constant volume
- Capacitance pressure gauge for the downstream side and a feed pressure transducer
- Gas manifold with automated valves for CO2, N2, CH4, H2, O2 and He
- Dry vacuum pump for evacuation between gases
- Controller with touch-screen HMI, automatic sequences, leak-rate check and time-lag calculation
Applications: Universities and R&D centres developing CO2-selective polymers, mixed-matrix membranes and thin-film composites; membrane makers screening new materials; materials scientists separating diffusion and sorption effects.
Send your membranes, gases and test conditions to mail@techincresearch.com or call +1 (289) 622-1032 for a P&ID, GA drawing and quotation.