top of page

The Tech Inc. Sampling Cabinet for H2S and Moisture Analysers (model TI-SCS-HS) conditions high-pressure natural gas for on-line hydrogen sulphide (H2S) and moisture measurement. It lowers the sample pressure, keeps the gas warm and removes liquids and particles before the analysers. Both measurements are easily disturbed: H2S is reactive and toxic, and moisture readings shift with any condensation in the sample path. The cabinet keeps the sample representative from the pipeline to the analyser.

Download the datasheet (PDF)

 

Main components

  • Inlet isolation valve and particle filter
  • Heated pressure regulator
  • Membrane separator and coalescer for liquid removal
  • Flow control with pressure gauges and flowmeters
  • Stainless steel sample path, with inert-treated surfaces to client specification
  • Calibration gas connection and vent to a safe location
  • Insulated stainless steel cabinet with thermostat-controlled heater
  • H2S gas detection at the cabinet, to client specification

 

Applications: sour natural gas, sales gas H2S and moisture limits, gas sweetening and dehydration unit outlets, transmission pipelines and metering stations, and gas compression plants.

 

Send your process data and specification to mail@techincresearch.com or call +1 (289) 622-1032 for a P&ID, GA drawing and quotation.

Sampling Cabinet for H2S and Moisture Analysers

bottom of page