top of page

The Tech Inc. Hollow Fibre Module Integrity Tester checks finished hollow-fibre modules for broken fibres and leaking tube sheets before they leave the factory. The tester wets the module, pressurises one side with air and measures pressure decay or diffusion flow against the module limit.

Download the datasheet (PDF)

 

Main components

  • Module holders for several diameters and lengths
  • Air pressure regulation and precision transducer
  • Pressure decay, bubble point and diffusion tests
  • Water fill and drain for wetting
  • Touch screen with pass/fail limits and reports

 

Applications: Hollow-fibre UF and MF module manufacturing and repair, and site checks of installed modules.

 

Send your module size, fibre and resin details and target output to mail@techincresearch.com or call +1 (289) 622-1032 for a layout, datasheet and quotation.